Dimension Vista LOCI Estradiol Test Code E2 Catalog Number: K6463 SMN: 10489099

Siemens Healthcare Diagnostics Inc. / initiated 2016-01-13 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc began a recall of Dimension Vista LOCI Estradiol Test Code E2 Catalog Number: K6463 SMN: 10489099 Product Usage: The E2 method is an in vitro diagnostic test for the... on . FDA classified it as Class II and gives the reason as "The drug fulvestrant (Faslodex) may cause falsely elevated estradiol results". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1241-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista LOCI Estradiol Test Code E2 Catalog Number: K6463 SMN: 10489099 Product Usage: The E2 method is an in vitro diagnostic test for the quantitative measurement of estradiol in human serum and plasma on the Dimension VistaSystem. Estradiol measurements are used in the diagnosis and treatment of various hormonal sexual disorders and in assessing placental function in complicated pregnancy.
Product code
CHP Radioimmunoassay, Estradiol
Reason for recall
The drug fulvestrant (Faslodex) may cause falsely elevated estradiol results.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens issued an Urgent Medical Device Correction (UMDC) to affected Siemens Healthcare Diagnostics customers in the United States on January 13, 2016 to be delivered to customers on January 14, 2016 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 13, 2016. The letters identified affected product, problem and the actions to be taken. Customers may continue to use the estradiol assays to report patient results however, the assays should not be used for patients being treated with fulvestrant. Customers where instructed to complete and return the Correction Effectiveness Check attached. For questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Quantity in commerce
1945 units
Distribution
Worldwide Distribution US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1241-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1240-2016Class IITerminatedVoluntary: Firm initiated
Z-1241-2016Class IITerminatedVoluntary: Firm initiated
Z-1242-2016Class IITerminatedVoluntary: Firm initiated
Z-1243-2016Class IITerminatedVoluntary: Firm initiated