Z-1242-2016Class II
IMMULITE 1000 Estradiol Test Code E2 Catalog Number: LKE21, LKE21(D), LKE25 SMN: 10381132, 10702832, 10381142
Siemens Healthcare Diagnostics Inc. / initiated 2016-01-13 / Terminated / root cause: design
Siemens Healthcare Diagnostics, Inc began a recall of IMMULITE 1000 Estradiol Test Code E2 Catalog Number: LKE21, LKE21(D), LKE25 SMN: 10381132, 10702832, 10381142 Product Usage: For in vitro diagnostic use with... on . FDA classified it as Class II and gives the reason as "The drug fulvestrant (Faslodex) may cause falsely elevated estradiol results". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1242-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- IMMULITE 1000 Estradiol Test Code E2 Catalog Number: LKE21, LKE21(D), LKE25 SMN: 10381132, 10702832, 10381142 Product Usage: For in vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers - for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various hormonal sexual disorders
- Product code
- CHP Radioimmunoassay, Estradiol
- Reason for recall
- The drug fulvestrant (Faslodex) may cause falsely elevated estradiol results.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens issued an Urgent Medical Device Correction (UMDC) to affected Siemens Healthcare Diagnostics customers in the United States on January 13, 2016 to be delivered to customers on January 14, 2016 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 13, 2016. The letters identified affected product, problem and the actions to be taken. Customers may continue to use the estradiol assays to report patient results however, the assays should not be used for patients being treated with fulvestrant. Customers where instructed to complete and return the Correction Effectiveness Check attached. For questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
- Quantity in commerce
- 9251 units
- Distribution
- Worldwide Distribution US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1240-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1241-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1242-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1243-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
