Fresenius Naturalyte Liquid Acid Concentrate Catalog Number: 08-2301-3
Fresenius Medical Care Renal Therapies Group, LLC began a recall of Fresenius Naturalyte Liquid Acid Concentrate Catalog Number: 08-2301-3 Intended Use: Acid Concentrate for Bicarbonate Dialysis on . FDA classified it as Class II and gives the reason as "On the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of constituents represents the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0887-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Fresenius Naturalyte Liquid Acid Concentrate Catalog Number: 08-2301-3 Intended Use: Acid Concentrate for Bicarbonate Dialysis
- Product code
- KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Reason for recall
- On the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of constituents represents the actual calcium concentration of the final dialysate, 3.0 mEq/L.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- The firm, Fresenius Medical Care, issued an "URGENT-MEDICAL DEVICE RECALL" letter dated 1/14/2015 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine your stock for the product; if found, discontinue use immediately; place all units in a secure, segregated area; contact your FMCNA Customer Service Team at 1-800-323-5188 for instructions how to return the affected lot (lot# 15STAC072), and promptly complete and return the attached reply form by Fax to: 781-699-9796 Attn: Quality Department or scan and email to: [email protected]. Fresenius will pick up and replace your product. If you have any additional medical concerns or questions, please contact Medical Information and Communication at 855-816-2309 or Website: www.fresenius-medinfo.com. Further information or support concerning this issue, contact the Customer Service Care Team at 800-323-5188 and reference the and Support Field Action Number.
- Quantity in commerce
- 848 (212 cases)
- Distribution
- US Distribution to states of: KS, LA, MS, OK, and TX.
- Product codes and lots
- Lot Number: 15STAC072 Exp Date: 2017-12-31
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0887-2016 | Class II | Terminated | Voluntary: Firm initiated |