Kopans Breast Lesion Localization Needle The Kopans Breast Lesion Localization Needle consists of a bevel-tip...
Cook Inc. began a recall of Kopans Breast Lesion Localization Needle The Kopans Breast Lesion Localization Needle consists of a bevel-tip needle with centimeter markings and a hookwire... on . FDA classified it as Class II and gives the reason as "Product potentially exposed to body fluids during processing. Potential adverse events that may occur as a result of the foreign matter include blood borne pathogen transmission. However, the risk of transmission of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1028-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Kopans Breast Lesion Localization Needle The Kopans Breast Lesion Localization Needle consists of a bevel-tip needle with centimeter markings and a hookwire. The hookwire features a burnish mark that allows the operator to ensure that the hookwire is within the needle tip during needle manipulation. The Kopans Breast Lesion Localization Needle is intended for preoperative marking of nonpalpable breast lesions.
- Product code
- MIJ Needle, Tumor Localization
- Reason for recall
- Product potentially exposed to body fluids during processing. Potential adverse events that may occur as a result of the foreign matter include blood borne pathogen transmission. However, the risk of transmission of blood borne pathogens is thought to be minimal to non-existent since these devices were sterilized after the potential exposure to body fluids.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Cook Medical sent an URGENT: MEDICAL DEVICE RECALL dated January 8, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to do the following: Please review the attached list of affected products and lot numbers shipped to your account, and quarantine any affected product that remains unused. 2. Immediately collect and return all unused affected products to Cook Medical as soon as possible for credit. 3. Please complete the attached Recall Response Form and return to Cook Medical either with the product or separately. 4. Please report any Adverse Event to Cook Medical Customer Relations 1-800- 457-4500 or 1-812-339-2235. Monday through Friday between 7:30 a.m. and 5:00 p.m. Eastern Daylight Time or email at [email protected] Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235.
- Quantity in commerce
- 134
- Distribution
- Nationwide distribution including AZ, CA, GA, IL, MO, MS, MT, NC, NH, NY, OH, OK, and PA..
- Product codes and lots
- Catalog # DKBL-20-7.0-A Lot # 6215428 date of manufacture 09/17/2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1028-2016 | Class II | Terminated | Voluntary: Firm initiated |