Z-0823-2016Class II

This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the...

Leica Microsystems, Inc. / initiated 2015-12-30 / Terminated / root cause: labeling and UDI

Leica Microsystems, Inc. began a recall of This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of... on . FDA classified it as Class II and gives the reason as "The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining when using a detection protocol where one of its components - Peroxide Block / Peroxidase Block is applied...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0823-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of human Pax-5 in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using the automated Bond system.
Product code
NJT Immunohistochemistry Reagents And Kits
Reason for recall
The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining when using a detection protocol where one of its components - Peroxide Block / Peroxidase Block is applied after the primary marker incubation as specified in the Instructions for Use Following the Instructions for Use.
Root cause tag
labeling and UDI
FDA root cause
Component change control
Action
The firm sent out a Medical Device Recall letter dated December 29th 2015 to two-hundred (200) consignees in the United States decribing the recall and the neccesary actions to be taken. These included: - Do not use or continue to use the listed antibody with the specified detection kits. - Order replacement detection kits for use with the listed antibody. - There is no need for retrospective review of tissue staining if appropriate positive controls were properly used and interpreted during the immuno / in-situ staining. -pass this notice primarily to the end users where the product has been sold and to all those within your organization who need to be aware of this issue. -Confirm reciept of the Recall Notice letter by signing and dating the attached Acknowledgement Form and faxing it back to us at 1-847-236-3747, or scan and email it to [email protected] Customer are requested to contact their local Leica representative if you have any questions or concerns, or alternatively please contact us at the following email address: [email protected] Peter Lloyd Leica Biosystems Newcastle Ltd Balliol Business Park West Benton Lane Newcastle upon Tyne NE 12 8EW United Kingdom Telephone +44 191 215 0567 Facsimile: +44 191 215 1152
Quantity in commerce
1700 units
Distribution
US Consignees incluude: AR ,AZ ,CA ,CO ,CT ,DC ,DE ,FL ,GA ,HI ,IA ,IL ,KY ,LA ,MA ,MD ,MI ,MN ,MO ,MT ,NC ,NH ,NJ ,NY ,OH ,OK ,PA ,PR ,SC ,TN ,TX ,UT ,VA ,VT ,WA ,WI and WV. Foreign Consignees include:Argentina , Australia , Brazil , Bulgaria , Chile , China , Ecuador ,Egypt , El Salvador , Finland , France , Georgia , Germany , Greece , India , Italy , Japan , Jordan , Kuwait , Lebanon , Malaysia , Philippines , Poland , Portugal , Romania , Russian Fed. , Singapore , South Korea , Spain , Sweden , Switzerland , Thailand , Tunisia , Turkey , United Kingdom and United Arab Emirates.
Product codes and lots
Product Code/Name: PA0552 (All Lots) Bond Ready-to-Use Primary Antibody PAX-5 (1EW).   Detection Kit: DS9800 BondTM Polymer Refine Detection Kit. Affected Detection Kit Lots: 42314, 42388, 42474, 42486, 42499, 42555, 42570, 42579, 42627, 42645, 42683, 42780, 42802, 42823, 42852, 43050.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0823-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0823-2016Class IITerminatedVoluntary: Firm initiated