Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0817-2016Class II

MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator...

Nihon Kohden America Inc / initiated 2012-01-19 / Terminated / root cause: design

Nihon Kohden America Inc began a recall of MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator Box) JB-116B, JB-132B (Amplifiers) Electromyography... on . FDA classified it as Class II and gives the reason as "The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0817-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nihon Kohden America Inc
Product
MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator Box) JB-116B, JB-132B (Amplifiers) Electromyography. (EMG)
Product code
GWF Stimulator, Electrical, Evoked Response
Reason for recall
The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow.
Root cause tag
design
FDA root cause
Device Design
Action
A customer notification letter dated 9/25/12 was sent to customers informing them of a potential malfunction of NIHON KOHDEN amplifier units, models JB-116B and JB-132B and Extension Unit MS-110B, used with the MEE-1000A Neuromaster Neural Function Measuring System. The letter informs the customers of the potential malfunction and that the probability that the malfunction occurs is extremely rare. The letter informs the customers of the preventive measure and the preventive corrective action. Customers are instructed to inform all potential users and operators of the Neural Function Measuring System MEE-1000A Neuromaster about the Corrective Action notice. Customers are instructed that NIHON KOHDEN technical support representative will contact them to arrange an update of their amplifier products.
Quantity in commerce
477 units total (137 units in US)
Distribution
Worldwide distribution. US nationwide, Australia, Austria, France, Germany, Greece, Japan, Romania, South Africa, South Korea, Spain, and United Kingdom.
Product codes and lots
Serial Numbers: MS-110B: (17 units) 00013-00019, 00023- 00032 MS-120BK: (6 units) 0009- 00014 JB-116B: (31 units), 0008, 00012, 00014, 00016, 00018, 00019, 00020, 00021, 00023, 00025, 00026, 00027, 00028, 00029, 00031, 00032, 00034, 00035, 00039, 00043, 00102,00103, 00122-00128, 00134, 00135 JB-132B: (83 units) 00009, 00011, 00012, 00014, 00015, 00016, 00018, 00019, 00020, 00021, 00023, 00024, 00025, 00026, 00027, 00036, 00037, 00038, 00041, 00042, 00043, 00044, 00046, 00047, 00049, 00061, 00062, 00066, 00067, 00068, 00069,00071, 00072, 00074, 00075, 00082, 00084, 0085, 00086, 00090, 00095, 00099, 00120, 00121-00129, 00131-00147, 00149-00155, 00157, 00161, 00164-00168

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0817-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0817-2016Class IITerminatedVoluntary: Firm initiated