Z-0804-2016Class I

Flow-i Anesthesia System

Maquet Cardiovascular Us Sales, LLC / initiated 2015-12-10 / Terminated / root cause: manufacturing process

Maquet Cardiovascular Us Sales, Llc began a recall of Flow-i Anesthesia System Product Usage: The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire... on . FDA classified it as Class I and gives the reason as "Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0804-2016
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Flow-i Anesthesia System Product Usage: The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment by healthcare professionals trained in inhalation anesthesia administration.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues. Only the Model C20 is affected in the US.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Maquet sent "Urgent Medical Device Field Action Field Correction For Flow-i Anesthesia Systems" letters/Field Correction Response forms dated 12/10/2015 via UPS to affected customers. The notification informed customers of the issue, asked them to distribute the notification to appropriate staff; verify the serial number(s) and informed them they could continue to use the system if they were not experiencing electrical issues. Customers were informed that a Maquet service representative would be contacting them to arrange a time to replace their affected units. Customers were asked to complete and return the enclosed Field Correction Response Form to Maquet at [email protected] or by fax to 973.860.1675. If the customer had any related questions they were instructed to contact Maquet Technical Support at 1.888.627.8383 (option 3; option 1; option 1) Monday through Friday between 8:00 am and 5:00 pm EST.
Quantity in commerce
7 units within US
Distribution
US Nationwide in the states of CA, NY, and OR
Product codes and lots
Model numbers: C20, C30 and C40 (Only Model # C20 for US)  Part numbers: 66 77 200 (US Only), 66 77 300 and 66 77 400  Serial numbers (2743, 2852, 2854, 2855, 2856, 2858

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0804-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0804-2016Class ITerminatedVoluntary: Firm initiated