Z-0859-2016Class II

Specify 5-6-5 Lead, Specify 2x8 Lead, Model Numbers: 39565, 39286. Spinal cord stimulation leads. Part of a...

Medtronic Neuromodulation / initiated 2016-01-11 / Terminated / root cause: labeling and UDI

Medtronic Neuromodulation began a recall of Specify 5-6-5 Lead, Specify 2x8 Lead, Model Numbers: 39565, 39286. Spinal cord stimulation leads. Part of a neurostimulation system for treating chronic... on . FDA classified it as Class II and gives the reason as "Some Specify 5-6-5 and 2x8 surgical lead product labeling does not adequately distinguish between permanent surgical implant and trialing/screening use". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0859-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Specify 5-6-5 Lead, Specify 2x8 Lead, Model Numbers: 39565, 39286. Spinal cord stimulation leads. Part of a neurostimulation system for treating chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
Some Specify 5-6-5 and 2x8 surgical lead product labeling does not adequately distinguish between permanent surgical implant and trialing/screening use.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Consignees were sent a Medtronic "Medical Device Correction" letter dated January 2016. The letter was addressed to Healthcare Professional. The letter described the problem and the product involved in the recall. Advised consignees to ensure their staff of the action that is being recommended. Requested consignees to complete and return the "Physician Confirmation Form". For additional information they can contact Medtronic Neuromodulation Technical services at 1-800-707-0933 weekends 7 am - 6 Pm Central time.
Quantity in commerce
51,138 (44,006 US; 9,132 OUS)
Distribution
Worldwide Distribution, including US Nationwide.
Product codes and lots
All lot/serial numbers are affected for both Model numbers: 39565 and 39286

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0859-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0859-2016Class IITerminatedVoluntary: Firm initiated