Z-0886-2016Class II

Auto Endo5 Hem-o-lok Automatic Clip Applier, Medium Large Polymer Clips, Rx Only, Sterile, REF 543965

Teleflex Medical / initiated 2016-02-02 / Terminated / root cause: manufacturing process

Teleflex Medical began a recall of Auto Endo5 Hem-o-lok Automatic Clip Applier, Medium Large Polymer Clips, Rx Only, Sterile, REF 543965 Product Usage: The Weck Auto Endo5 hem-o-lok ML automatic... on . FDA classified it as Class II and gives the reason as "The clip in the applier may be missing a boss. A boss is the part of the clip that is used to hold the clip within the applier jaws prior to application". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0886-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
Auto Endo5 Hem-o-lok Automatic Clip Applier, Medium Large Polymer Clips, Rx Only, Sterile, REF 543965 Product Usage: The Weck Auto Endo5 hem-o-lok ML automatic endoscopic ligating clip appliers are indicated for use as delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips. These appliers are designed for use with 5/5.5mm cannulas. Hem-o-lok Ligating Clips are intended for use in procedures involving ligation of vessels or tissue structures.
Product code
FZP Clip, Implantable
Reason for recall
The clip in the applier may be missing a boss. A boss is the part of the clip that is used to hold the clip within the applier jaws prior to application.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Teleflex sent an Urgent Medical Device Recall Notification letter dated February 2, 2016 to customers. The letter identified the affected product, problem and actions to be taken. A separate letter was sent to distributors advising them to conduct a sub-recall. Customers were instructed to: 1. If you have affected stock, immediately discontinue use and quarantine any products with the catalog numbers listed in the letter. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1- 855-419-8507, Attn: Customer Service or email to [email protected] will document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will document receipt of the letter. For questions contact your local sales representative or Customer Service at 1-866-246-6990
Quantity in commerce
543 ea.
Distribution
Worldwide Distribution in the state of CA, MI, NC, PA, TX, VA, and the countries of Canada, Korea & Belgium
Product codes and lots
73J1500065, 73J1500066, 73J1500280, 73J1500440, 73J1500565, 73J1500566, 73K1500193

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0886-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0886-2016Class IITerminatedVoluntary: Firm initiated