Z-1045-2016Class II

Maquet Getinge Group 1180.19XO Extension Device. Manufacturer MAQUET GmbH, Used to support the upper and...

Maquet Cardiovascular Us Sales, LLC / initiated 2016-01-07 / Terminated / root cause: manufacturing process

Maquet Cardiovascular Us Sales, Llc began a recall of Maquet Getinge Group 1180.19XO Extension Device. Manufacturer MAQUET GmbH, Used to support the upper and lower extremities immediately before, during and after... on . FDA classified it as Class II and gives the reason as "MAQUET Medical Systems USA is initiating a field action for correction of the MAQUET Extension Device (Part No 1180.19F0). The Extension Device is manufactured by Maquet GmbH, in Germany, and is a table top module for...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1045-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Maquet Getinge Group 1180.19XO Extension Device. Manufacturer MAQUET GmbH, Used to support the upper and lower extremities immediately before, during and after fracture surgery. The Extension Device (1180.19F0) is a table top module for use with the Maquet MAGNUS operating table system and it is attached to the base mounting point of the MAGNUS operating table.
Product code
FQO Table, Operating-Room, Ac-Powered
Reason for recall
MAQUET Medical Systems USA is initiating a field action for correction of the MAQUET Extension Device (Part No 1180.19F0). The Extension Device is manufactured by Maquet GmbH, in Germany, and is a table top module for use with the Maquet MAGNUS operating table system. Maquet GmbH was made aware of a potential issue that can result in the Extension Device becoming unlocked from the table top, causing it to fall down.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Maquet sent "URGENT PRODUCT RECALL MEDICAL DEVICE CORRECTION" letters/MEDICAL DEVICE CORRECTION RESPONSE FORM dated January 7, 2016, via UPS to the affected customers. The notification informed customers of the issue with the affected product; how to identify affected product; measures to be taken by the users; and that a MAQUET service representative will be contacting them to schedule servicing of the device to install a Support Frame Lock. Customers with questions were asked to contact their local MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383 (Option 2, Option 3) Monday through Friday between 8:00 am and 5:00 pm EDT. Customers were also requested to fill out and return the attached response forms.
Quantity in commerce
2 within US; 247 OUS
Distribution
Worldwide Distribution - US including AZ and MO. Foreign distribution to Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, Ecuador, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, and Venezuela.
Product codes and lots
Catalog Number 1180.19F0; Serial numbers 00003 and 00004.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1045-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1045-2016Class IITerminatedVoluntary: Firm initiated