510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0813-2016Class II

Dual Stage Venous Return Catheter, Low-Profile, wire reinforced (36F/46F, Sterile), Model Number: LRD-61046...

LivaNova USA, Inc. / initiated 2016-01-25 / Terminated / root cause: manufacturing process

Sorin Group USA, Inc. began a recall of Dual Stage Venous Return Catheter, Low-Profile, wire reinforced (36F/46F, Sterile), Model Number: LRD-61046, Expiration 10/31/2018, Product is packed in a... on . FDA classified it as Class II and gives the reason as "Sorin Group initiated field correction for the Dual Stage Venous Return Catheter because the incorrect Instructions For Use (IFU) was packaged in the case". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0813-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Dual Stage Venous Return Catheter, Low-Profile, wire reinforced (36F/46F, Sterile), Model Number: LRD-61046, Expiration 10/31/2018, Product is packed in a sealed pouch, ten (10) pouches per case. Cardiopulmonary bypass catheter.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Sorin Group initiated field correction for the Dual Stage Venous Return Catheter because the incorrect Instructions For Use (IFU) was packaged in the case.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
LivaNova sent an Urgent Safety Alert dated January 25, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Advice on action to be taken: "Remove and discard the IFU included within the case packaging for the product. "A copy of the correct IFU (Part Number 435085202, Revision B) is attached with this letter "Print copies of the IFU and place with the product. For questions regarding this notice, please contact Customer Service at 1-800-650-2623 (Monday to Friday BAM to 5PM MT) or e-mail [email protected]. We sincerely apologize for this inconvenience and thank you in advance for your support
Quantity in commerce
60
Distribution
US Distribution to CA., and Internationally to Canada.
Product codes and lots
lot - 1529300018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0813-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0813-2016Class IITerminatedVoluntary: Firm initiated