Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation
Draeger Medical, Inc. began a recall of Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation on . FDA classified it as Class I and gives the reason as "The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0827-2016
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Draeger Medical, Inc. will send an Urgent Medical Device Recall Letter to consignees in a method that confirms delivery. For consignees that have received new devices with firmware version 1.50 (127 devices) or through device updates (644 devices; reference Recall Z-0436-2016), those devices will be downgraded to firmware version 1.49 and the batteries replaced. Batteries will continue to be replaced for all consignees every six months until a final solution is available.
- Quantity in commerce
- 2553
- Distribution
- US Nationwide Distribution including Puerto Rico.
- Product codes and lots
- Catalog #s 8416400 and 8417400
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0827-2016 | Class I | Terminated | Voluntary: Firm initiated |