Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0803-2016Class II

Olympus Scientific Solutions Americas Corporation ( OSSA ) Delta XRF Analyzer . This is a Analytical X-ray...

Evident Scientific MIS, Inc. / initiated 2016-01-13 / Terminated / root cause: software

Olympus Scientific Solutions Americas began a recall of Olympus Scientific Solutions Americas Corporation ( OSSA ) Delta XRF Analyzer . This is a Analytical X-ray system on . FDA classified it as Class II and gives the reason as "The Firm has discovered a Software bug". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0803-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Evident Scientific MIS, Inc.
Product
Olympus Scientific Solutions Americas Corporation ( OSSA ) Delta XRF Analyzer . This is a Analytical X-ray system
Product code
RBY X-Ray Fluorescence
Reason for recall
The Firm has discovered a Software bug.
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
Olympus Scientific Solutions Americas ( OSSSA) planned action to bring product into compliance: Notification will be given to purchasers per 21 CFR 1003.21. A software modification will be made available to all users contacted. The letter to affected purchasers makes clear that the product will be brought into compliance without charge and offers the purchaser several options for applying the software update. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Tracking of customer/owner self-installation(s) of the software update should be done to the maximum possible extent via serial-number- controlled downloads, follow-up email reporting, return-mail card or some other similarly effective means. For further questions you may call : (866) 446-6689.
Quantity in commerce
US - 5000 units
Distribution
Worldwide Distribution - USA (nationwide) Distribution
Product codes and lots
All models of the Delta XRF Analyzer

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0803-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0803-2016Class IITerminatedVoluntary: Firm initiated