Z-1079-2016Class II

Siemens Healthcare Diagnostic ADVIA Centaur Calibrator E- 6- Pack (Reference) Catalog Number: 04636889 SMN...

Siemens Healthcare Diagnostics Inc. / initiated 2016-02-04 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc began a recall of Siemens Healthcare Diagnostic ADVIA Centaur Calibrator E- 6- Pack (Reference) Catalog Number: 04636889 SMN: 10335532 ADVIA Centaur Calibrator E For in vitro... on . FDA classified it as Class II and gives the reason as "ADVIA Centaur Systems Calibrator E Lot 42 and 43 Bias with the Testosterone Assay". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1079-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Healthcare Diagnostic ADVIA Centaur Calibrator E- 6- Pack (Reference) Catalog Number: 04636889 SMN: 10335532 ADVIA Centaur Calibrator E For in vitro diagnostic use in calibration the following assays using ADVIA Centaur systems: Cortisol, Progesterone Testosterone. ADVIA Centaur Testosterone For in vitro diagnostic use in the quantitative determination of total testosterone {bound and unbound) in serum using the ADVIA Centaur and ADVIA Centaur XP Systems
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
ADVIA Centaur Systems Calibrator E Lot 42 and 43 Bias with the Testosterone Assay
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Siemens Healthcare issued an Urgent Medical Device Correction (UMDC) to all affected Siemens Healthcare Diagnostics customers in the United States on February 4, 2016 to be delivered to customers on February 5, 2016. An Urgent Field Safety Notice (UFSN) was sent out for distribution to customers outside the United States on February 4, 2016. These notices inform the customers of the lot to lot bias of Calibrator E with the ADVIA Centaur Testosterone assay and advise customers to move to Calibrator E kit lots ending in 44. Fax-back forms will be used to confirm that the customers received the Urgent Medical Device Correction. Questions, contact your Siemens Customer Care Center or your local Siemens technical support representative
Quantity in commerce
519 pack
Distribution
Worldwide Distribution US (nationwide) and Internationally to: Canada AR AU BD BR CA CH CL CN CO DO EC GT HK ID IE IL IN JP KR LK MM MX MY NI NP NZ PE PH PY SA SG TH TW UY VN ZA
Product codes and lots
Lot codes/Exp Date: 91005A42 May 9, 2016; 53698A42 May 9, 2016

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1079-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1077-2016Class IITerminatedVoluntary: Firm initiated
Z-1078-2016Class IITerminatedVoluntary: Firm initiated
Z-1079-2016Class IITerminatedVoluntary: Firm initiated