Z-1122-2016Class II

Product 24 consists of all product under

Zimmer Manufacturing B.V. / initiated 2016-01-11 / Terminated / root cause: packaging

Zimmer Manufacturing B.V. began a recall of Product 24 consists of all product under product code: HRS and same usage: Item no: 234703104 DISTAL MEDIAL FEMORAL CON 234703206 DISTAL MEDIAL FEMORAL CON... on . FDA classified it as Class II and gives the reason as "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1122-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer Manufacturing B.V.
Product
Product 24 consists of all product under product code: HRS and same usage: Item no: 234703104 DISTAL MEDIAL FEMORAL CON 234703206 DISTAL MEDIAL FEMORAL CON 234703208 DISTAL MEDIAL FEMORAL CON 234800504 DISTAL POSTERIOR/LATERAL 234800506 DISTAL POSTERIOR/LATERAL 234800508 DISTAL POSTERIOR/LATERAL 234800510 DISTAL POSTERIOR/LATERAL 234800604 DISTAL POSTERIOR/LATERAL 234800606 DISTAL POSTERIOR/LATERAL 234800608 DISTAL POSTERIOR/LATERAL 234800610 DISTAL POSTERIOR/LATERAL 234800614 DISTAL POSTERIOR/LATERAL Product Usage: Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Product code
HRS Plate, Fixation, Bone
Reason for recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root cause tag
packaging
FDA root cause
Packaging
Action
Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Quantity in commerce
37
Distribution
US Nationwide Distribution
Product codes and lots
lot no.: 61983647 62296609 62152979 62056816 61771916 61932061 61771919 62056817 61771920 61866489 61880386 61859716 61906564 62056818 61866490 61880387 61880388 61895979 61895980

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1122-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1099-2016Class IITerminatedVoluntary: Firm initiated
Z-1100-2016Class IITerminatedVoluntary: Firm initiated
Z-1101-2016Class IITerminatedVoluntary: Firm initiated
Z-1102-2016Class IITerminatedVoluntary: Firm initiated
Z-1103-2016Class IITerminatedVoluntary: Firm initiated
Z-1104-2016Class IITerminatedVoluntary: Firm initiated
Z-1105-2016Class IITerminatedVoluntary: Firm initiated
Z-1106-2016Class IITerminatedVoluntary: Firm initiated
Z-1107-2016Class IITerminatedVoluntary: Firm initiated
Z-1108-2016Class IITerminatedVoluntary: Firm initiated
Z-1109-2016Class IITerminatedVoluntary: Firm initiated
Z-1110-2016Class IITerminatedVoluntary: Firm initiated
Z-1111-2016Class IITerminatedVoluntary: Firm initiated
Z-1112-2016Class IITerminatedVoluntary: Firm initiated
Z-1113-2016Class IITerminatedVoluntary: Firm initiated
Z-1114-2016Class IITerminatedVoluntary: Firm initiated
Z-1115-2016Class IITerminatedVoluntary: Firm initiated
Z-1116-2016Class IITerminatedVoluntary: Firm initiated
Z-1117-2016Class IITerminatedVoluntary: Firm initiated
Z-1118-2016Class IITerminatedVoluntary: Firm initiated
Z-1119-2016Class IITerminatedVoluntary: Firm initiated
Z-1120-2016Class IITerminatedVoluntary: Firm initiated
Z-1121-2016Class IITerminatedVoluntary: Firm initiated
Z-1122-2016Class IITerminatedVoluntary: Firm initiated
Z-1123-2016Class IITerminatedVoluntary: Firm initiated
Z-1124-2016Class IITerminatedVoluntary: Firm initiated
Z-1125-2016Class IITerminatedVoluntary: Firm initiated
Z-1126-2016Class IITerminatedVoluntary: Firm initiated
Z-1127-2016Class IITerminatedVoluntary: Firm initiated
Z-1128-2016Class IITerminatedVoluntary: Firm initiated
Z-1129-2016Class IITerminatedVoluntary: Firm initiated
Z-1130-2016Class IITerminatedVoluntary: Firm initiated
Z-1131-2016Class IITerminatedVoluntary: Firm initiated
Z-1132-2016Class IITerminatedVoluntary: Firm initiated
Z-1133-2016Class IITerminatedVoluntary: Firm initiated
Z-1134-2016Class IITerminatedVoluntary: Firm initiated
Z-1135-2016Class IITerminatedVoluntary: Firm initiated
Z-1136-2016Class IITerminatedVoluntary: Firm initiated
Z-1137-2016Class IITerminatedVoluntary: Firm initiated
Z-1138-2016Class IITerminatedVoluntary: Firm initiated
Z-1139-2016Class IITerminatedVoluntary: Firm initiated
Z-1140-2016Class IITerminatedVoluntary: Firm initiated
Z-1141-2016Class IITerminatedVoluntary: Firm initiated
Z-1142-2016Class IITerminatedVoluntary: Firm initiated
Z-1143-2016Class IITerminatedVoluntary: Firm initiated
Z-1144-2016Class IITerminatedVoluntary: Firm initiated
Z-1145-2016Class IITerminatedVoluntary: Firm initiated
Z-1146-2016Class IITerminatedVoluntary: Firm initiated
Z-1147-2016Class IITerminatedVoluntary: Firm initiated
Z-1148-2016Class IITerminatedVoluntary: Firm initiated
Z-1149-2016Class IITerminatedVoluntary: Firm initiated
Z-1150-2016Class IITerminatedVoluntary: Firm initiated
Z-1151-2016Class IITerminatedVoluntary: Firm initiated
Z-1152-2016Class IITerminatedVoluntary: Firm initiated
Z-1153-2016Class IITerminatedVoluntary: Firm initiated
Z-1154-2016Class IITerminatedVoluntary: Firm initiated
Z-1155-2016Class IITerminatedVoluntary: Firm initiated
Z-1156-2016Class IITerminatedVoluntary: Firm initiated
Z-1157-2016Class IITerminatedVoluntary: Firm initiated
Z-1158-2016Class IITerminatedVoluntary: Firm initiated
Z-1159-2016Class IITerminatedVoluntary: Firm initiated
Z-1160-2016Class IITerminatedVoluntary: Firm initiated
Z-1161-2016Class IITerminatedVoluntary: Firm initiated
Z-1162-2016Class IITerminatedVoluntary: Firm initiated
Z-1163-2016Class IITerminatedVoluntary: Firm initiated
Z-1164-2016Class IITerminatedVoluntary: Firm initiated
Z-1165-2016Class IITerminatedVoluntary: Firm initiated
Z-1166-2016Class IITerminatedVoluntary: Firm initiated
Z-1167-2016Class IITerminatedVoluntary: Firm initiated
Z-1168-2016Class IITerminatedVoluntary: Firm initiated
Z-1169-2016Class IITerminatedVoluntary: Firm initiated
Z-1170-2016Class IITerminatedVoluntary: Firm initiated