Z-1146-2016Class II

Product 48 consists of all product under

Zimmer Manufacturing B.V. / initiated 2016-01-11 / Terminated / root cause: packaging

Zimmer Manufacturing B.V. began a recall of Product 48 consists of all product under product code: JWH and same usage: Item no: 597206526 NEXGEN ALL-POLY PATELLA, for use in total knee arthroplasty on . FDA classified it as Class II and gives the reason as "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1146-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer Manufacturing B.V.
Product
Product 48 consists of all product under product code: JWH and same usage: Item no: 597206526 NEXGEN ALL-POLY PATELLA, for use in total knee arthroplasty
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root cause tag
packaging
FDA root cause
Packaging
Action
Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Quantity in commerce
0
Distribution
US Nationwide Distribution
Product codes and lots
lot no.: 61873094 61895696 61905916 61905918 61905920 61919766 61919767 61919768 61956249 61956250 61981202 61981204 61999375 62027331 62060784 62074467 62074469 62096750 62096751 62133740 62164984 62168517 62171693 62171694 62171696 62174936 62205101 62213332 62213333 62220943 62345691 62345692 62345693 62345694 62352447 62352450

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1146-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1099-2016Class IITerminatedVoluntary: Firm initiated
Z-1100-2016Class IITerminatedVoluntary: Firm initiated
Z-1101-2016Class IITerminatedVoluntary: Firm initiated
Z-1102-2016Class IITerminatedVoluntary: Firm initiated
Z-1103-2016Class IITerminatedVoluntary: Firm initiated
Z-1104-2016Class IITerminatedVoluntary: Firm initiated
Z-1105-2016Class IITerminatedVoluntary: Firm initiated
Z-1106-2016Class IITerminatedVoluntary: Firm initiated
Z-1107-2016Class IITerminatedVoluntary: Firm initiated
Z-1108-2016Class IITerminatedVoluntary: Firm initiated
Z-1109-2016Class IITerminatedVoluntary: Firm initiated
Z-1110-2016Class IITerminatedVoluntary: Firm initiated
Z-1111-2016Class IITerminatedVoluntary: Firm initiated
Z-1112-2016Class IITerminatedVoluntary: Firm initiated
Z-1113-2016Class IITerminatedVoluntary: Firm initiated
Z-1114-2016Class IITerminatedVoluntary: Firm initiated
Z-1115-2016Class IITerminatedVoluntary: Firm initiated
Z-1116-2016Class IITerminatedVoluntary: Firm initiated
Z-1117-2016Class IITerminatedVoluntary: Firm initiated
Z-1118-2016Class IITerminatedVoluntary: Firm initiated
Z-1119-2016Class IITerminatedVoluntary: Firm initiated
Z-1120-2016Class IITerminatedVoluntary: Firm initiated
Z-1121-2016Class IITerminatedVoluntary: Firm initiated
Z-1122-2016Class IITerminatedVoluntary: Firm initiated
Z-1123-2016Class IITerminatedVoluntary: Firm initiated
Z-1124-2016Class IITerminatedVoluntary: Firm initiated
Z-1125-2016Class IITerminatedVoluntary: Firm initiated
Z-1126-2016Class IITerminatedVoluntary: Firm initiated
Z-1127-2016Class IITerminatedVoluntary: Firm initiated
Z-1128-2016Class IITerminatedVoluntary: Firm initiated
Z-1129-2016Class IITerminatedVoluntary: Firm initiated
Z-1130-2016Class IITerminatedVoluntary: Firm initiated
Z-1131-2016Class IITerminatedVoluntary: Firm initiated
Z-1132-2016Class IITerminatedVoluntary: Firm initiated
Z-1133-2016Class IITerminatedVoluntary: Firm initiated
Z-1134-2016Class IITerminatedVoluntary: Firm initiated
Z-1135-2016Class IITerminatedVoluntary: Firm initiated
Z-1136-2016Class IITerminatedVoluntary: Firm initiated
Z-1137-2016Class IITerminatedVoluntary: Firm initiated
Z-1138-2016Class IITerminatedVoluntary: Firm initiated
Z-1139-2016Class IITerminatedVoluntary: Firm initiated
Z-1140-2016Class IITerminatedVoluntary: Firm initiated
Z-1141-2016Class IITerminatedVoluntary: Firm initiated
Z-1142-2016Class IITerminatedVoluntary: Firm initiated
Z-1143-2016Class IITerminatedVoluntary: Firm initiated
Z-1144-2016Class IITerminatedVoluntary: Firm initiated
Z-1145-2016Class IITerminatedVoluntary: Firm initiated
Z-1146-2016Class IITerminatedVoluntary: Firm initiated
Z-1147-2016Class IITerminatedVoluntary: Firm initiated
Z-1148-2016Class IITerminatedVoluntary: Firm initiated
Z-1149-2016Class IITerminatedVoluntary: Firm initiated
Z-1150-2016Class IITerminatedVoluntary: Firm initiated
Z-1151-2016Class IITerminatedVoluntary: Firm initiated
Z-1152-2016Class IITerminatedVoluntary: Firm initiated
Z-1153-2016Class IITerminatedVoluntary: Firm initiated
Z-1154-2016Class IITerminatedVoluntary: Firm initiated
Z-1155-2016Class IITerminatedVoluntary: Firm initiated
Z-1156-2016Class IITerminatedVoluntary: Firm initiated
Z-1157-2016Class IITerminatedVoluntary: Firm initiated
Z-1158-2016Class IITerminatedVoluntary: Firm initiated
Z-1159-2016Class IITerminatedVoluntary: Firm initiated
Z-1160-2016Class IITerminatedVoluntary: Firm initiated
Z-1161-2016Class IITerminatedVoluntary: Firm initiated
Z-1162-2016Class IITerminatedVoluntary: Firm initiated
Z-1163-2016Class IITerminatedVoluntary: Firm initiated
Z-1164-2016Class IITerminatedVoluntary: Firm initiated
Z-1165-2016Class IITerminatedVoluntary: Firm initiated
Z-1166-2016Class IITerminatedVoluntary: Firm initiated
Z-1167-2016Class IITerminatedVoluntary: Firm initiated
Z-1168-2016Class IITerminatedVoluntary: Firm initiated
Z-1169-2016Class IITerminatedVoluntary: Firm initiated
Z-1170-2016Class IITerminatedVoluntary: Firm initiated