M/L Taper Hip Prosthesis (Item: 00-7711-012-20, Lot: 62937093

Zimmer GmbH / initiated 2016-02-11 / Terminated / root cause: labeling and UDI

Zimmer Biomet, Inc. began a recall of M/L Taper Hip Prosthesis (Item: 00-7711-012-20, Lot: 62937093 on . FDA classified it as Class II and gives the reason as "Zimmer Biomet initiated a lot-specific voluntary recall on 02/11/2016, of the M/L Taper Hip Prosthesis. single manufacturing lot of Size 12.5, Extended Offset, M/L Taper hip stems was incorrectly etched. The device was...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1046-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
M/L Taper Hip Prosthesis (Item: 00-7711-012-20, Lot: 62937093
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
Zimmer Biomet initiated a lot-specific voluntary recall on 02/11/2016, of the M/L Taper Hip Prosthesis. single manufacturing lot of Size 12.5, Extended Offset, M/L Taper hip stems was incorrectly etched. The device was etched as STD, rather than EXT. The product packaging is labeled correctly as Extended. Offset, and the device is an extended offset stem.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Zimmer, Inc. initiated a lot-specific voluntary recall on 02/11/2016, of the M/L Taper Hip Prosthesis. Electronic notices were e-mailed to distributors, and letters were mailed via certified mail on 02/11/2016 to distributors and hospitals. A field complaint investigation confirmed that this single manufacturing lot of Size 12.5, Extended Offset, M/L Taper hip stems was incorrectly etched. The device was etched as STD, rather than EXT. The product packaging is labeled correctly as Extended Offset, and the device is an extended offset stem. Distributors will return on-hand product to Zimmer Biomet and ensure all of their products are accounted for using the form provided in the letter. Hospital risk managers and surgeons will be provided with a letter identifying the issue and their responsibilities. These responsibilities include: Assisting the Zimmer Biomet sales representative with the quarantine of the product and completion and return of the Acknowledgement Form. Customers are asked to complete the Certificate of Acknowledgement Form and return to [email protected]. Customers with have further questions or concerns please call the customer call center at 1-800-348-2759 between 8:00 am and 8:00 pm EST.
Quantity in commerce
9 units
Distribution
Distributed in the states of SD, MO, NY, MI, PA, KY and TN and in the country of Argentina.
Product codes and lots
Item: 00-7711-012-20, Lot: 62937093

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1046-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1046-2016Class IITerminatedVoluntary: Firm initiated