Z-1088-2016Class II

Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray...

Medtronic Navigation, Inc / initiated 2016-01-29 / Terminated / root cause: manufacturing process

Medtronic Navigation, Inc. began a recall of Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging... on . FDA classified it as Class II and gives the reason as "Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1088-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Medtronic Navigation issued a Field Safety Notice dated January 29, 2016, to notify them of a hardware correction in the unit. A Medtronic service representative will service the system and arrange for an inspection of the system. This representative will properly torque the eight (8) screws if they find them to be loose. In the meantime, if noise or vibration are observed during use of the system customers should contact their Medtronic representative to report the issue. Customers with questions were instructed to call 720-890-3160. For questions regarding this recall call 978-698-6000.
Quantity in commerce
1 unit
Distribution
Worldwide Distribution to Austria only.
Product codes and lots
Serial Number: C0936

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1088-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1088-2016Class IITerminatedVoluntary: Firm initiated