Z-1239-2016Class II
Alaris PC Unit, Infusion Pump Model 8000, Part No. TC10005092
CareFusion 303, Inc / initiated 2016-02-12 / Terminated
CareFusion 303, Inc. began a recall of Alaris PC Unit, Infusion Pump Model 8000, Part No. TC10005092 on . FDA classified it as Class II and gives the reason as "CareFusion is recalling the Alaris PC unit because a component on the PC unit power supply may cause a "System Error" or "Missing Battery" error code (120.4630)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1239-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris PC Unit, Infusion Pump Model 8000, Part No. TC10005092
- Product code
- FRO Dressing, Wound, Drug
- Reason for recall
- CareFusion is recalling the Alaris PC unit because a component on the PC unit power supply may cause a "System Error" or "Missing Battery" error code (120.4630).
- FDA root cause
- Under Investigation by firm
- Action
- A customer notification letter dated 2/25/16 will be sent to all their customers who purchased the Alaris PC unit (Model 8000) to inform them that CareFusion is recalling it because a component on the PC unit power supply board may cause a "System Error" or "Missing Battery" error code (120.4630). The letter informs the customers of the potential risk involved and the actions to be taken. Customers with recall related questions are instructed to call CareFusion Support Center at (888) 562-6018, 7am-4pm PST, Mon-Fri or email at [email protected]. Customers with clinical inquires, product complaints or clinical troubleshooting are instructed to contact customer advocacy at (888) 812-3266, 24 hrs/day, 7 days/wk or email at [email protected]. Customers with technical questions regarding the Alaris system are instructed to contact technical support at (888) 812-3229, 6am-5pm PST, Mon-Fri or email at [email protected].
- Quantity in commerce
- 170 units
- Distribution
- Nationwide in US: AR, AZ, CA, FL, GA, IN, KY, MD, MI, MN, MO, NJ, NM, NY, OH, OR, PA, SC, TN, TX, VA, WI.
- Product codes and lots
- Serial No. 4044845 9950535 4064411 12475162 4045659 9899143 3606478 4021398 4093501 4043745 4063430 4094994 9907675 9945292 9922622 4043375 12409069 4020663 12498854 4044531 4058193 12351849 4064030 9925068 12426138 4063830 4063911 9945269