Product 55 consists of all product under
Zimmer Manufacturing B.V. began a recall of Product 55 consists of all product under product code: HSB and same usage: Item no: 225900132 ONE-PIECE LAG SCREW 11MM 225900135 ONE-PIECE LAG SCREW 11MM... on . FDA classified it as Class II and gives the reason as "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1153-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer Manufacturing B.V.
- Product
- Product 55 consists of all product under product code: HSB and same usage: Item no: 225900132 ONE-PIECE LAG SCREW 11MM 225900135 ONE-PIECE LAG SCREW 11MM 225900137 ONE-PIECE LAG SCREW 11MM 225900140 ONE-PIECE LAG SCREW 11MM 225900142 ONE-PIECE LAG SCREW 11MM 225900145 ONE-PIECE LAG SCREW 11MM 225900152 ONE-PIECE LAG SCREW 11MM Usage: An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
- Quantity in commerce
- 58320
- Distribution
- US Nationwide Distribution
- Product codes and lots
- lot no.: 62133636 61882374 61964347 61987304 62133643 62171285 62298481 61969229 62080367 62292385 61657722 61930479 61930480 61930481 62109268 61136016 61930482 62098917 62098918 61764507 61764508 61764509 61895897 61895905 61895907 61895908
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1099-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1100-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1101-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1102-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1103-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1104-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1105-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1106-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1107-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1108-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1109-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1110-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1111-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1112-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1113-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1114-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1115-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1116-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1117-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1118-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1119-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1120-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1121-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1122-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1123-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1124-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1125-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1126-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1127-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1128-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1129-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1130-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1131-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1132-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1133-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1134-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1135-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1136-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1137-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1138-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1139-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1140-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1141-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1142-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1143-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1144-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1145-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1146-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1147-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1148-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1149-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1150-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1151-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1152-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1153-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1154-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1155-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1156-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1157-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1158-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1159-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1160-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1161-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1162-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1163-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1164-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1165-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1166-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1167-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1168-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1169-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1170-2016 | Class II | Terminated | Voluntary: Firm initiated |