510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1195-2016Class II

Arthrex IBalance TKA Tibial Tray Size 2 - Catalog # AR-503-TTTC, Size 3 - Catalog # AR-503-TTTD, Size 4...

Arthrex GmbH / initiated 2015-12-17 / Terminated / root cause: manufacturing process

Arthrex, Inc. began a recall of Arthrex IBalance TKA Tibial Tray Size 2 - Catalog # AR-503-TTTC, Size 3 - Catalog # AR-503-TTTD, Size 4 -Catalog # AR-503-TTTE, Size 5 - Catalog # AR-503-TTTF... on . FDA classified it as Class II and gives the reason as "Devices reported to have a smooth texture to the outer surface of the metal, which differed from a previous that had a textured outer surface". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1195-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arthrex GmbH
Product
Arthrex IBalance TKA Tibial Tray Size 2 - Catalog # AR-503-TTTC, Size 3 - Catalog # AR-503-TTTD, Size 4 -Catalog # AR-503-TTTE, Size 5 - Catalog # AR-503-TTTF, Size 6 - Catalog # AR-503-TTTG , Size 7 -Catalog # AR-503-TTTH, and Size 8 - Catalog # AR-503-TTTJ. Intended for implantation with bone cement, with the exception of the porous coated femoral components which can be used cemented or uncemented (biological fixation).
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Devices reported to have a smooth texture to the outer surface of the metal, which differed from a previous that had a textured outer surface.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On February 5, 2016 Arthrex sent Urgent Medical Device Voluntary Recall letters to their distributors and Surgeons/Medical Facilities reporting the removal. The letter identified the reason for the recall, the potential patient impact, and actions to be taken. Users were to discontinue use of the identified affected devices and follow the instructions provided to return the affected devices that have not been implanted. To return the devices, contact Arthrex at 1-866-267-9138 or by fax at 1-239-591-6943 or via e-mail at [email protected]. Customers must complete and return the Acknowledgement and Receipt.
Quantity in commerce
2,378 units.
Distribution
Nationwide Distribution -- AR, AZ, CA, CO, DC, DE, FL, GA, IL, IN, KS, MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, WA, WI and District of Columbia.
Product codes and lots
Size 2 Unique Device Identifier: 00888867106703 Batch #'s: 927692, 915207, 982771, 979834, 982774, 1206164, 1206165, 1206167, 1206166, 1227388, 1227387, 1257195, 1233620, 1289648, 1233621, 1289653, 1289736, 1289654, 1336430, 1336431, 1305721, 1306332, 1352062, 1336433, 1336436, 1352058, 1352063, 1376104, 1376106, and 1376107.  Size 3 Unique Device Identifier: 00888867106710 Batch #'s: 780838, 952789, 1206169, 1206170, 1206168, 1191337, 1191337, 1272728, 1297048, 1297044, 1297049, 1272730, 1272729, 1307262, 1297060, 1297057, 1297365, 1307263, 1336445, 1307267, 1336451, 1336448, 1336451, 1307265, 1336453, 1346958, 1346962, 1346982, 1346994, 1346965, 1336452, 1379085, 1379104, 1379110, 1379113, 1379114, 1346999, and 1379084.  Size 4 Unique Device Identifier: 00888867106727 Batch #'s: 894339, 913623, 874733, 876482, 876478, 1191338, 1206172, 1206173, 1297075, 1297084, 1315459, 1297086, 1297366, 13115453, 1315454, 1315459, 1315462, 1315464, 1336466, 1336465, 1336467, 1336455, 1336468, 13336469, 1356527, 1356530, 1356531, 1356529, 1356574, 1379205, 1379203, 1379204, 1379206, 1379200, 1356576, 1379196, and 1379201. Size 5 Unique Device Identifier: 00888867106734 Batch #'s:780840, 971202, 1206176, 1206177, 1356586, 1322523, 1380216, 1380223, 1380224, 1356588, 1380221, 1380226, 1555653, and 1555654. Size 6 Unique Device Identifier: 00888867106741 Batch #'s:780841, 970368, 1206178, 1206179, 1322459, 1322460, 1322524, 1357785, 1380243, 1380244, 1380245, 1380246, 1357786, and 1380242. Size 7 Unique Device Identifier: 00888867106758 Batch #'s: 780842, 982767, 984546, 1206180, 1337230, 1206181, 1337234, 1337237, 1337240, 1337241, 1337256, 1358297, 1358299, 1358298, 1358300, 1358302, 1380708, 1380709, and 1380710. Size 8 Unique Device Identifier: 00888867106765 Batch #'s: 913624, 970865, 957056, 1191342, 1224471, 120800, 1191341, 1191339, 876484, 876485, 1228632, 1251021, 1292154, 1251022, 1281699, 1292161, 1293756, 1293758, 1307294, 1307295, 1308538, 1337262, 1337300, 1337303, 1347012, 1347013, 1337325, 1347014, 1380743, 1347025, 1380747, 1380748, and 1380742.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1195-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1195-2016Class IITerminatedVoluntary: Firm initiated