Z-1253-2016Class II

10mm X 12o Solitaire Ti Medium Spacer Solitaire Anterior Spinal System, Model Number 1400-1230

Highridge Medical, LLC / initiated 2016-02-12 / Terminated / root cause: labeling and UDI

Biomet Spine, LLC began a recall of 10mm X 12o Solitaire Ti Medium Spacer Solitaire Anterior Spinal System, Model Number 1400-1230. Product Usage: The SOLITAIRETM SPINAL SYSTEM is designed for... on . FDA classified it as Class II and gives the reason as "Zimmer Biomet Spine initiated a recall of certain Solitaire Ti Spacers because they may have a mislabeled lordotic angle of 12degrees when the angle is actually 6 degrees". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1253-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Highridge Medical, LLC
Product
10mm X 12o Solitaire Ti Medium Spacer Solitaire Anterior Spinal System, Model Number 1400-1230. Product Usage: The SOLITAIRETM SPINAL SYSTEM is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Additionally, the SOLITAIRETM ANTERIOR SPINAL SYSTEM is indicated for use in the thoracolumbar spine (i.e., T10 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The SolitaireTM System is also indicated for treating fractures of the thoracic and lumbar spine. The SolitaireTM System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.
Product code
OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar
Reason for recall
Zimmer Biomet Spine initiated a recall of certain Solitaire Ti Spacers because they may have a mislabeled lordotic angle of 12degrees when the angle is actually 6 degrees.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
Zimmer Biomet sent an Urgent Medical Device Recall Notice letter dated February 26, 2016 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the enclosed Response Form. For questions call 303-465-8929.
Quantity in commerce
10
Distribution
Worldwide Distribution - US Nationwide - one foreign consignee (not Canada).
Product codes and lots
P/N1400-1230 from Lot#2350871 from either loose inventory or Kit 555000162, Date of Manufacture: April 16, 2013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1253-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1253-2016Class IITerminatedVoluntary: Firm initiated