Z-1247-2016Class II

Neurotherm Curved Radiofrequency (RF) cannnula 10 cm 18 gauge, 10 mm active tip Model Number: C-1010-R-18

Neurotherm, Inc. / initiated 2013-02-26 / Terminated / root cause: manufacturing process

Neurotherm, Inc. began a recall of Neurotherm Curved Radiofrequency (RF) cannnula 10 cm 18 gauge, 10 mm active tip Model Number: C-1010-R-18 Product Usage: The device is a radiofrequency needle... on . FDA classified it as Class II and gives the reason as "Straight needle labeled as a curved needle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1247-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Neurotherm, Inc.
Product
Neurotherm Curved Radiofrequency (RF) cannnula 10 cm 18 gauge, 10 mm active tip Model Number: C-1010-R-18 Product Usage: The device is a radiofrequency needle, used for the application of radiofrequency energy with as intended result, thermo coagulation of nervous tissue in the human body for pain relief. They are designed to be used to treat chronic pain. During nerve ablation, the needle is used in conjunction with a radiofrequency generator, an electrode and grounding pads.
Product code
GXI Probe, Radiofrequency Lesion
Reason for recall
Straight needle labeled as a curved needle
Root cause tag
manufacturing process
FDA root cause
Process control
Action
NeuroTherm notified consignees by telephone between February 2- March 15, 2013 about the affected product, problem and actions to be taken. Customers were instructed to examine inventory and return devices.
Quantity in commerce
135 cannulas (27 boxes/5)
Distribution
US Nationwide Distribution in the states of CA ,CO, OH, OR
Product codes and lots
Lot number: 12131-1C

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1247-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1247-2016Class IITerminatedVoluntary: Firm initiated