Z-1098-2016Class II
Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a...
IMRIS Imaging, Inc. / initiated 2016-02-10 / Terminated / root cause: design
Deerfield Imaging began a recall of Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD) on . FDA classified it as Class II and gives the reason as "Emergency helium venting lines (i.e. quench lines) may not have been properly installed". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1098-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- IMRIS Imaging, Inc.
- Product
- Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Emergency helium venting lines (i.e. quench lines) may not have been properly installed.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were sent a IMRIS "Urgent Field Safety Notice" letter dated February 10, 2016. The letter described the problem and the product involved in the recall. The firm also stated that it is important that customers review the portion of the quench line vent described and ensure that it is installed to the specifications supplied as part of the Planning Guide. Customers were reminded that it is important to have an emergency plan in place in the event Helium gas escapes into the magnet room or any other area. For questions customers can contact their local customer support hotline or +1-866-475-0525.
- Quantity in commerce
- 35 USA, 14 OUS
- Distribution
- Worldwide Distribution -- US, including the states of CA, CT, FL, GA, IL, IN, MD, MA, MN, MO, NH, NC, OH, OR, PA, TX, UT, VA, and WI; and ,countries of CHINA, CANADA, AUSTRALIA, FRANCE, JAPAN, and GERMANY.
- Product codes and lots
- 70036, 10076-21, M00004, 10073-01, 23529, 40481, M00152, M00005, 30445, M00169, 060014, 10043-00, 30823, 10004680, 10050-01, 40662, 30935, 31084, 31203, 10002221, 40510, 40613, 30719, 31046, 30910, 30832, 30623, 31213, 10004152, 10072-01, 31250, 40815, 31049, 30068, 40412, M0038, M00106, M00367, 10035-01, 40475, 40495, 10007498, 10005946, 40263, 40688, 10058-00, 40619, 31085, and 10005639.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1098-2016 | Class II | Terminated | Voluntary: Firm initiated |