Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1394-2016Class II

VERION Reference Unit (Vision Planner)

Alcon Laboratories / initiated 2016-03-01 / Terminated / root cause: software

Alcon Research, Ltd. began a recall of VERION Reference Unit (Vision Planner) Product Usage: The VERION Image Guided system is an advanced surgical planning, imaging and guidance technology designed... on . FDA classified it as Class II and gives the reason as "Alcon is conducting a voluntary medical device correction of all VERION Reference Units (Vision Planner) that are shared with the Alcon LenSx Laser System after receiving reports concerning the inclusion of unplanned...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1394-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alcon Laboratories
Product
VERION Reference Unit (Vision Planner) Product Usage: The VERION Image Guided system is an advanced surgical planning, imaging and guidance technology designed to provide greater accuracy and efficiency during cataract surgery. It allows eye surgeons to capture a high-resolution, diagnostic reference image of the patients eye pre-operatively, quickly determine an optimized surgical plan that enables surgeons to see all inclusions and alignment in real-time.
Product code
NFG Device, Communications, Images, Ophthalmic
Reason for recall
Alcon is conducting a voluntary medical device correction of all VERION Reference Units (Vision Planner) that are shared with the Alcon LenSx Laser System after receiving reports concerning the inclusion of unplanned arcuates on printed, saved (.pdf) or exported surgical plans.
Root cause tag
software
FDA root cause
Software change control
Action
Alcon sent an Urgent Medical Device Correction letter dated February 29, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that an Alcon Technical Services Engineers will contact customers, schedule Service Calls, and perform onsite corrections. For questions call 1-800-862-5266.
Quantity in commerce
97 units
Distribution
Worldwide Distribution - US Nationwide throughout the US and the countries of: Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Czech Republic, Chile, China, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Greece, Guatlemala, Hungary, India, Indoesia, Isreal, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Nicaragua, Norway, Panama, Philippines, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, UK, Ukrain, United Arab Emirates
Product codes and lots
Serial Numbers: 1017, 1119, 1123, 1125, 1126, 1128, 1133, 1135, 1136, 1137, 1138, 1139, 1142, 1147, 1150, 1153, 1155, 1161, 1165, 1168, 1172, 1178, 1181, 1182, 1183, 1188, 1190, 1191, 1193, 1196, 1197, 1198, 1199, 1201, 1203, 1206, 1207, 1220, 1226, 1228, 1230, 1232, 1250, 1256, 1258, 1259, 1421, 1433, 1435, 1439, 1451, 1452, 1506, 1508, 1509, 1523, 1536, 1548, 1559, 1562, 1565, 1568, 1581, 1597, 1606, 1628, 1632, 1649, 1661, 1663, 1668, 1671, 1694, 1696, 1705, 1716, 1726, 1729, 1798, 1816, 1818, 1820, 1824, 1829, 1921, 1922, 1934, 1937, 1938, 1940, 1942, 1954, 1957, 2057, 2069, 2086, 2092

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1394-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1394-2016Class IITerminatedVoluntary: Firm initiated