Z-1267-2016Class II

The Spectra-System Dental Implant 2008 system is comprised of dental implant fixtures and prosthetic devices...

Implant Direct Sybron Manufacturing LLC / initiated 2016-03-04 / Terminated / root cause: manufacturing process

Implant Direct Sybron Manufacturing, LLC began a recall of The Spectra-System Dental Implant 2008 system is comprised of dental implant fixtures and prosthetic devices that compose a two-piece implant system. GoDirect... on . FDA classified it as Class II and gives the reason as "Implant Direct Sybron Manufacturing, LLC is recalling GoDirect Implants because some consignees did not receive the correct tool to drive the implant to bone level". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1267-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Implant Direct Sybron Manufacturing LLC
Product
The Spectra-System Dental Implant 2008 system is comprised of dental implant fixtures and prosthetic devices that compose a two-piece implant system. GoDirect Implants Part Number: 423011, 423013, 424708, 424710, 424713,433010,433710
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
Implant Direct Sybron Manufacturing, LLC is recalling GoDirect Implants because some consignees did not receive the correct tool to drive the implant to bone level.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Implant Direct sent an Urgent: Medical Device Field Corrective Action letter dated March 4, 2016, to all affected customers. The letter states that Implant Direct will correct the issue and send the proper tool. The proper tool is included with the notification. The letter requests that customers complete and return the Acknowledgement and Recall Return Form within 48 hours. It customers are an authorized Implant Direct Sybron Manufacturing distributor it is requested that customers are identified and contact them to inform them of the issue within 48 hours of receiving the notification. Distributors should also provide customers with the correct tool. If you have any of the affected product listed above and have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425. For further questions regarding this please call (818) 444-3300 Ext. 3323
Quantity in commerce
2033
Distribution
Worldwide Distribution - Nationwide Distribution to the states of : FL, AZ, IL, MS, TX, CA, ID, AL, MO, MN, NY, CO, GA, ME, NJ, NC, NV, LA, KS, MN, KY, PA, UT, OH, MI, SC, MD, WA, WI., and to the countries of : Canada, Honduras and Dominican Republic.
Product codes and lots
Lot Number: 29707, 29708, 29709, 32444, 37438, 38159, 39506, 40394, 40542, 44661, 46030, 50644, 50824, 51092, 51249, 52798, 54274, 54275, 54617, 55407,60158

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1267-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1267-2016Class IITerminatedVoluntary: Firm initiated