Z-1501-2016Class II

CareLink iPro2 Reports, Model No. MMT-7340 is software intended to retrospectively calibrates sensor data and...

Medtronic, Inc. / initiated 2016-02-19 / Terminated / root cause: manufacturing process

Medtronic MiniMed Inc. began a recall of CareLink iPro2 Reports, Model No. MMT-7340 is software intended to retrospectively calibrates sensor data and provides reports of continuous glucose information on . FDA classified it as Class II and gives the reason as "Medtronic MiniMed is recalling 19 CareLink iPro Therapy Management reports because the CareLink iPro reports were created using the Pacific Standard Time zone as the time reference, although the patients and their...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1501-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
CareLink iPro2 Reports, Model No. MMT-7340 is software intended to retrospectively calibrates sensor data and provides reports of continuous glucose information
Product code
OUG Medical Device Data System
Reason for recall
Medtronic MiniMed is recalling 19 CareLink iPro Therapy Management reports because the CareLink iPro reports were created using the Pacific Standard Time zone as the time reference, although the patients and their sensor data were subject to the Central Standard Time zone. The CareLink iPro2 Professional CGM (continuous glucose monitor) data uploading error that resulted in an inaccurate time zone being applied to 19 CareLink iPro reports.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
An Urgent Field Safety Notice was sent to customers on 2/19/16 to make customers aware of an iPro2 Professional CGM (continuous glucose monitor) data uploading error that resulted in an inaccurate time zone being applied to 21 CareLink iPro reports. The letter informs customers that the affected CareLink iPro reports were created using the Pacific Standard Time zone as the time reference, although the patients and their sensor data were subject to the Central Standard Time zone. The letter informs the customers of the requested next steps and to contact Maneesh Shrivastav, Director of Strategy at Medtronics Diabetes by email at [email protected] or call (763) 526-3509 if they have any questions.
Quantity in commerce
20
Distribution
US: MN
Product codes and lots
n/a

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1501-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1501-2016Class IITerminatedVoluntary: Firm initiated