Z-1395-2016Class II
BARD ArcticGel Pads
C.R. Bard Inc / initiated 2016-03-04 / Terminated / root cause: manufacturing process
C.R. Bard, Inc. began a recall of BARD ArcticGel Pads Product Usage: The Arctic Sun Temperature Management System is intended for monitoring and controlling patient temperature. ArcticGel Pads... on . FDA classified it as Class II and gives the reason as "Complaints were received of restricted/inaccurate flow rate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1395-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- C.R. Bard Inc
- Product
- BARD ArcticGel Pads Product Usage: The Arctic Sun Temperature Management System is intended for monitoring and controlling patient temperature. ArcticGel Pads are only for use with an Arctic Sun Temperature Management System Control Module.
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- Complaints were received of restricted/inaccurate flow rate.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Bard Medical sent an Urgent Medical Product Recall letter dated March 4, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to not use any of the affected products and to return any unused product Bard Medical Division via the provided Customer Recall and Effectiveness Form. BMD will be providing a replacement for the returned product. If the product was further distributed, the consignee was instructed to identify their customers and notify them of the product recall.
- Quantity in commerce
- 21,412 units
- Distribution
- Worldwide Distribution - Domestic: US Nationwide; Foreign: Argentina, Australia, Austria, Belarus, Belgium, Bosnia, Brazil Canada, Chile, China, Colombia, Croatia, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Oman Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, , South Korea, Kuwait, Macedonia, Malaysia, Mexico, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, Turkey, United Arab Emirates, Ukraine, United Kingdom, Uruguay and Vietnam.
- Product codes and lots
- All lots with Catalog numbers: 31700 (Universal Pad), 3170004 (Universal Pad 4 pk), 31703 (XSmall pad kit), 3170302 (XSmall pad kit 2 pk), 31705 (Small pad kit), 3170502 (Small pad kit 2 pk), 31707 (Medium pad kit), 3170702 (Medium pad kit 2 pk), 31709 (Large pad kit) and 3170902 (Large pad kit 2 pk), that are within their current expiration dates.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1395-2016 | Class II | Terminated | Voluntary: Firm initiated |