Helios HL3T and HLT Track Mounted Dental Lights
Pelton & Crane / Marus / DCI Equipment / KaVo began a recall of Helios HL3T and HLT Track Mounted Dental Lights Product Usage: The product is intended for use by dental professionals to illuminate the oral cavity while... on . FDA classified it as Class II and gives the reason as "The firm received two customer complaints local Pelton & Crane distributors that the Helios 3000 track mounted dental lights were missing the two 8/32x3/16 set screws during the initial installation of the dental light". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1268-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- KaVo Dental Technologies, LLC
- Product
- Helios HL3T and HLT Track Mounted Dental Lights Product Usage: The product is intended for use by dental professionals to illuminate the oral cavity while performing examinations and dental procedures.
- Product code
- EAZ Light, Operating, Dental
- Reason for recall
- The firm received two customer complaints local Pelton & Crane distributors that the Helios 3000 track mounted dental lights were missing the two 8/32x3/16 set screws during the initial installation of the dental light.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Pelton & Crane sent an Urgent Medical Device Recall letters to affected customers on March 8, 2016 and will be completed by 3/11/2016. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the Field Correction Acknowledgement Form in the enclosed self-address stamped envelope to Pelton & Crane. For questions or concerns regarding this recall please call 1-800-659-6560, ext. 7227.
- Quantity in commerce
- 156 units
- Distribution
- US Nationwide Distribution in the states of AZ, AR, CA, CO, CT, FL, IA, ID, IL, IN, MA, MI, MN, NJ, NY, NC, OH, OR, PA, TN, TX, UT, VA, WI and Canada
- Product codes and lots
- Model/Serial No. - HLT-5535, 5543, 5536, 5537, 5538, 5556, 5540, 5541, 5545, 5546, 5515, 5500, 5503, 5520, 5521, 5547, 5548, 5549, 5550, 5551, 5552, 5510, 5558, 5522, 5523, 5524, 5525, 5526, 5527, 5528, 5529, 5530, 5531, 5532, 5501, 5502, 5504, 5505, 5534, 5508, 5509, 5533, 5516, 5517, 5518, 5519, 5544, 5553, 5554, 5555, 5542, 5557, 5506, 5507, 5511, 5498, 5539, 5499, 5512, 5513, 5514; Model/Serial No. HL3T-11110, 11102, 11103, 11104, 11105, 11106, 11099, 11164, 11165, 11166, 11167, 11193, 11170, 11171, 11118, 11119, 11120, 11145, 11146, 11147, 11148, 11149, 11121, 11123, 11150, 11113, 11114, 11115, 11116, 11117, 11168, 11169, 11158, 11187, 11133, 11144, 11136, 11137, 11138, 11139, 11140, 11141, 11142, 11143, 11177, 11111, 11112, 11135, 11190, 11191, 11192, 11100, 11182, 11183, 11185, 11186, 11134, 11122, 11124, 11125. 11126, 11127, 11128, 11129, 11130, 11131, 11132, 11155, 11156, 11157, 11159, 11160, 11161, 11107, 11108, 11176, 11172, 11173, 11174, 11175, 11109. 11151, 11152, 11153, 11154, 11101, 11181, 11184, 11188, 11189, 11178, 11179, 11180, 11162, 11163
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1268-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.