A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate...
Siemens Medical Solutions USA, Inc began a recall of A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical... on . FDA classified it as Class II and gives the reason as "To inform users about the possible incorrect values for Distance Measurements when using certain modalities in combination with syngo Imaging". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1349-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images, including digital mammography images.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- To inform users about the possible incorrect values for Distance Measurements when using certain modalities in combination with syngo Imaging.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens mailed a Customer Advisory Notice to their customers instructing them to check their modalities for the usage of the mentioned DICOM attributes with different values. They recommended upgrading the workplace(s) to syngo Imaging Studio version VB36D if they have such modalities and it cannot be ensured by the modality provider that only one attribute is filled or both attributes are filled with the same value. Customers were asked to contact their local service organization to clarify the compatibility of their syngo Imaging server version with workplace version VB36D.
- Quantity in commerce
- 49 imaging systems
- Distribution
- US (nationwide) including the states of AL, CA, CO, CT, FL, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OK, OR, PA, TN, TX, and WI.
- Product codes and lots
- Model # 10014063
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1349-2016 | Class II | Terminated | Voluntary: Firm initiated |