Z-1349-2016Class II

A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate...

Siemens Medical Solutions USA, Inc. / initiated 2016-01-18 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical... on . FDA classified it as Class II and gives the reason as "To inform users about the possible incorrect values for Distance Measurements when using certain modalities in combination with syngo Imaging". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1349-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images, including digital mammography images.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
To inform users about the possible incorrect values for Distance Measurements when using certain modalities in combination with syngo Imaging.
Root cause tag
software
FDA root cause
Software design
Action
Siemens mailed a Customer Advisory Notice to their customers instructing them to check their modalities for the usage of the mentioned DICOM attributes with different values. They recommended upgrading the workplace(s) to syngo Imaging Studio version VB36D if they have such modalities and it cannot be ensured by the modality provider that only one attribute is filled or both attributes are filled with the same value. Customers were asked to contact their local service organization to clarify the compatibility of their syngo Imaging server version with workplace version VB36D.
Quantity in commerce
49 imaging systems
Distribution
US (nationwide) including the states of AL, CA, CO, CT, FL, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OK, OR, PA, TN, TX, and WI.
Product codes and lots
Model # 10014063

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1349-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1349-2016Class IITerminatedVoluntary: Firm initiated