Siemens branded SOMATOM Emotion and SOMATOM Sensation families of CT systems intended to produce...
Siemens Medical Solutions USA, Inc began a recall of Siemens branded SOMATOM Emotion and SOMATOM Sensation families of CT systems intended to produce cross-sectional images of the body by computer reconstruction... on . FDA classified it as Class II and gives the reason as "Siemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1301-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens branded SOMATOM Emotion and SOMATOM Sensation families of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Siemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio licenses "Cardio Pro" and "Cardio Retro".
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A Safety Advisory Notice was sent to customers informing them of a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems. Siemens informed them that they will deliver an Update Instruction for their Service Engineers to modify the installed Firmware. This will correct the rare synchronization errors of Scanner Firmware and Scanner Software to start the application of the x-ray in compliance with the optimum triggering point for best image quality.
- Quantity in commerce
- 124 units
- Distribution
- US (nationwide) including Guam and the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, VA, and WA.
- Product codes and lots
- Model Number: SOMATOM Emotion 16 - 7734713
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1300-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1301-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1302-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1303-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1304-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1305-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1306-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1307-2016 | Class II | Terminated | Voluntary: Firm initiated |