Z-1232-2016Class II

syngo X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization...

Siemens Medical Solutions USA, Inc. / initiated 2016-02-24 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of syngo X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization, communication, and storage of medical images and data... on . FDA classified it as Class II and gives the reason as "After importing, the segmentation results appear mirrored at the CARTO system and can't be used for the ablation procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1232-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
syngo X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and post processing during interventional procedures. .
Product code
LLZ System, Image Processing, Radiological
Reason for recall
After importing, the segmentation results appear mirrored at the CARTO system and can't be used for the ablation procedure.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent an Important Customer Safety Notice dated February 24, 2016, to all affected customers notifying them that the issue will be remedied with a software update via AX004/16/S. Following the installation of this update, the segmentation results can be used at the CARTO system as described in the operator manual. Customers with questions were instructed to call 800-888-7436. For questions regarding this recall call 610-448-6461.
Quantity in commerce
6
Distribution
Nationwide Distribution to IL, NY, MT, and MN.
Product codes and lots
Serial # 20119, 20098, 20041, 20124, 12177, 20026

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1232-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1232-2016Class IITerminatedVoluntary: Firm initiated