Reveal LINQ insertable Cardiac Monitor, Model LNQ11. Monitors subcutaneous ECG continuously
Medtronic Inc. began a recall of Reveal LINQ insertable Cardiac Monitor, Model LNQ11. Monitors subcutaneous ECG continuously on . FDA classified it as Class II and gives the reason as "Medtronic has identified an issue with the sensitivity of an algorithm used in the Reveal LINQ ICM that may prematurely trigger the Recommended Replacement Time (RRT) alert in some devices. Battery capacity is not...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1249-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Reveal LINQ insertable Cardiac Monitor, Model LNQ11. Monitors subcutaneous ECG continuously.
- Product code
- DSI Detector And Alarm, Arrhythmia
- Reason for recall
- Medtronic has identified an issue with the sensitivity of an algorithm used in the Reveal LINQ ICM that may prematurely trigger the Recommended Replacement Time (RRT) alert in some devices. Battery capacity is not affected and the device will continue to support data collection and manual data transmissions. As part of the normal behavior of the device, 30 days after RRT status is reached, Reveal
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees in US were hand delivered a Medtronic "Medical Device Correction" letter dated February 2016. The letter was addressed to Physician or Healthcare Professional. The letter described the Issue & Patient Management Guidance. For questions contact your local Medtronic Representative or Medtronic Diagnostic Technical Services at 800-929-4043. Outside the U.S., the letter will be mailed or hand delivered to impacted consignees.
- Quantity in commerce
- 189,009
- Distribution
- Worldwide Distribution-US (nationwide) including DC and the countries of Andorra, Australia, Austria, Belgium, Botswana, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Faroe Islands , Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates and United Kingdom.
- Product codes and lots
- All Reveal LINQ ICM, Model LNQ11 serial numbers are impacted. Medtronic continues to manufacture and distribute Reveal Ll NQ devices.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1249-2016 | Class II | Terminated | Voluntary: Firm initiated |