Z-1308-2016Class II

VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear...

Mitsubishi Heavy Industries, Ltd., Hiroshima Machinery Work / initiated 2016-03-07 / Terminated / root cause: software

MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK began a recall of VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of... on . FDA classified it as Class II and gives the reason as "In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1308-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mitsubishi Heavy Industries, Ltd., Hiroshima Machinery Work
Product
VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is incorrectly set and if nobody is aware of it, an erroneous treatment would be given to the patient.
Root cause tag
software
FDA root cause
Software design
Action
A Field Safety Notice dated March 2, 2016, was sent to all affected customers via e-mail on March 7, 2016. The Field Safety Notice informed users of the problem and provided a safe workaround until corrected software is available. The correction will be free of charge and a Customer Service Representative will contact all affected customers to schedule its installation. Customers with questions were advised to contact their Customer Service Representative.
Quantity in commerce
25 Worldwide, US: 4 units
Distribution
Nationwide Distribution including New York, Texas, Florida, and Ohio
Product codes and lots
Software version 3.5.8, 3.5.2 and lower; Serial Numbers: 201902, 203901, 203919, 203924

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1308-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1308-2016Class IITerminatedVoluntary: Firm initiated