Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1530-2016Class II

Acorn 180 Curved Stairlift

Acorn Stairlifts, Inc / initiated 2016-02-17 / Terminated

Acorn Stairlifts Inc began a recall of Acorn 180 Curved Stairlift on . FDA classified it as Class II and gives the reason as "Aluminum rivets holding the base squab to the framer under prolonged customer use were seen to elongate and subsequently break". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1530-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Acorn Stairlifts, Inc
Product
Acorn 180 Curved Stairlift
Product code
PCD Powered Stairway Chair Lift
Reason for recall
Aluminum rivets holding the base squab to the framer under prolonged customer use were seen to elongate and subsequently break.
FDA root cause
Component change control
Action
Customers were notified by phone calls, documents mailed with a Rationalization Action plan to follow. Contact information 1-888-563-0410 x1500 or 1400
Quantity in commerce
1,290 USA and 9,607 Foreign
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AZ, CA, CO, DE, FL, GA, IL, IN, KS, KY, LA, MA, ME, MD, MI, MO, NH, NJ, NY, NC, OH, PA, RI, SC, TN, TX, UT, VT, VI, WA, WV, WI, and District of Columbia and to the countries of UK and Germany.
Product codes and lots
Serial Numbers: 140401487591 to 140410249946. All models manufactured before April 2012

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1530-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1530-2016Class IITerminatedVoluntary: Firm initiated