TOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider...
ArthroCare Corporation began a recall of TOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider 0.8 mm REF Q6000-01; TOPAZ EPF MicroDebrider 0.8... on . FDA classified it as Class II and gives the reason as "During functional testing of devices after real time aging, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1321-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ArthroCare Corporation
- Product
- TOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider 0.8 mm REF Q6000-01; TOPAZ EPF MicroDebrider 0.8 mm 45 REF Q6002-01
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- During functional testing of devices after real time aging, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Consignees were notified via letter on 3/3/2016. Consignees were instructed to quarantine the recalled products and complete an Inventory Return Certification Form.
- Quantity in commerce
- 26,903 devices
- Distribution
- Worldwide Distribution-US (nationwide) and the following countries-Austria, Belgium, Bulgaria, Canada, Denmark, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, China, Hong Kong, India, Lebanon, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arabic Emirates, Vietnam, Australia, Japan, New Zealand, Brazil, and Chile.
- Product codes and lots
- 1031506, 1031507, 1032123, 1032325, 1032741, 1032742, 1033443, 1033965, 1033966, 1034397, 1039120, 1041765, 1044089, 1044090, 1045862, 1048237, 1048933, 1049185, 1052071, 1056212, 1060586, 1060979, 1063394, 1064599, 1067696, 1067748, 1069174, 1071358, 1074125, 1075714, 1070521, 1070812, 1071974, 1078025, 1077499, 1082811, 1084095, 1084094, 1085458, 1086436, 1086437, 1087917, 1094313, 1097573, 1103824, 1107183, 1107297, 1113562, 1113563, 1114534, 1114535, 1114536, 1115896, 1031518, 1032308, 1032309, 1032310, 1032755, 1032756, 1033972, 1033973, 1034225, 1034226, 1035283, 1035284, 1038547, 1043507, 1046556, 1047064, 1048722, 1049266, 1051812, 1052070, 1054443, 1055730, 1059865, 1063661, 1067691, 1067273, 1067747, 1068019, 1069172, 1071357, 1072853, 1072854, 1070497, 1070808, 1071971, 1076849, 1077450, 1079140, 1079220, 1078115, 1081753, 1082361, 1082810, 1086380, 1094312, 1094312, 1094721, 1097571, 1101041, 1107295, 1109045, 1111935, 1111936, 1113534, 1113535, 1113536, 1114532, 1115888, 1115891, 1116959, 1116130, 1116720, 1117209, 1117214, 1031341, 1032311, 1033761, 1034446, 1035691, 1035692, 1041775, 1043962, 1046557, 1048236, 1050067, 1052651, 1057008, 1058808, 1059867, 1063665, 1064594, 1067695, 1068021, 1071319, 1071320, 1073441, 1074876, 1077498, 1080520, 1114533, 1115893, 1031340, 1032962, 1033266, 1045359, 1047065, 1051331, 1056849, 1064592, 1067694, 1070709
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1321-2016 | Class II | Terminated | Voluntary: Firm initiated |