Z-1281-2016Class II

CIOS ALPHA (Image-intensified fluoroscope x-ray system)

Siemens Medical Solutions USA, Inc. / initiated 2016-02-16 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the... on . FDA classified it as Class II and gives the reason as "Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1281-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens sent a Customer Safety Advisory Notice dated February 18, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers that Siemens is planning to replace the affected parts as a precautionary measure. A Siemens service engineer will contact each customer to schedule implementation of this precautionary measure.
Quantity in commerce
4 units
Distribution
US Nationwide Distributed in the states of AR, CA, NE and UT
Product codes and lots
Model Number: 10308191

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1281-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1281-2016Class IITerminatedVoluntary: Firm initiated