Z-1299-2016Class II

system, x-ray, angiographic, model # 10094135, 10094137, 10094139, 10094141, 10280959 Artis zee/ zeego is a...

Siemens Medical Solutions USA, Inc. / initiated 2016-02-26 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of system, x-ray, angiographic, model # 10094135, 10094137, 10094139, 10094141, 10280959 Artis zee/ zeego is a family of dedicated angiography systems developed... on . FDA classified it as Class II and gives the reason as "In Artis zeego systems, angulations in the vicinity of the C-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. In Artis systems with A100Plus generators, an attempt to...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1299-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
system, x-ray, angiographic, model # 10094135, 10094137, 10094139, 10094141, 10280959 Artis zee/ zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
In Artis zeego systems, angulations in the vicinity of the C-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. In Artis systems with A100Plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the X-ray tube) can result in the failure of a module in the high-voltage generator.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Siemens, sent an "Important Customer Safety Notice" dated 2/16/2016 to its customers. The notice informed them of a corrective action that will be performed to prevent a possible hazard to patients, operators, other persons or equipment. The notice described the product, problem and actions to be taken. Siemens plans to contact each customer to arrange a date to add additional protection mechanisms by means of a system software update. If you have any questions, contact Engineer, Regulatory Technical Specialist at 610-448-6461 or email: [email protected].
Quantity in commerce
552 units
Distribution
US Nationwide Distribution including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and PUERTO RICO.
Product codes and lots
Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1299-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1299-2016Class IITerminatedVoluntary: Firm initiated