Z-1274-2016Class I

Simplexa Group A Strep Direct (MOL2850) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model...

DiaSorin Molecular LLC / initiated 2016-02-10 / Terminated

Focus Diagnostics Inc began a recall of Simplexa Group A Strep Direct (MOL2850) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2850, MOL1451, MOL1452, MOL1455 on . FDA classified it as Class I and gives the reason as "Focus Diagnostics is recalling the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850) because...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1274-2016
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DiaSorin Molecular LLC
Product
Simplexa Group A Strep Direct (MOL2850) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2850, MOL1451, MOL1452, MOL1455
Product code
PGX Groups A, C And G Beta-Hemolytic Streptococcus Nucleic Acid Amplification System
Reason for recall
Focus Diagnostics is recalling the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850) because it may result in a false negative result, a false positive result or Error code (s) 500 or 505, and it may result in higher invalid rates.
FDA root cause
Under Investigation by firm
Action
A customer notification letter dated 2/10/16 was sent to all their customers who received the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850). The letter informs the customers that Focus Diagnostics is recalling MOL2150 and MOL2650 kits because it may result in a false negative result, a false positive result or Error Code(s) 500 or 505. Additionally, some customers are reporting insufficient specimen volume errors which may result in higher invalid rates. In those cases, the error results in a Report Note Statement of Insufficient Specimen Volume. The letter informs the customers of the recommendations to be taken for the identified kits and the risks involved for positive results, negative results, and delayed results. The letter informs the customers of the actions to be taken. Customers are instructed to sign the achnowledgement form and email the form to [email protected] or fax back to Focus Diagnostics Technical Services at (562) 240-6526 within 10 business days. On 2/11/16, Focus Diagnostics sent to their customers an amendment to the customer notification letter to clarify what is considered an "early Ct", Focus would consider Internal Control (IC) Ct's less than 25 as "early Ct". Customers are informed that if an internal control Ct is less than 25, additional testing may be necessary. On 2/15/16, Focus Diagnostics sent to their customers an amendment to the customer notification letter to inform them to determine if they had instances of early Cts less than 25, then to please refer to Section III Analyzing Runs, pages 6-10 and 6-15 in the Focus Diagnostics, Operator Manual for Integrated Cycler Studio 6.0 For Use with IVD Assays in the USA. Also, for a more detailed analysis, customers are instructed to view excerpts from the ICS Operator Manual covering Section III, pages 6-10 and 6-16. On 3/17/16,
Quantity in commerce
41 units
Distribution
Worldwide Distribution - US Nationwide in the states of: TX, CA, MA, NY, WA, MI, ME, IN, NJ, WV, OH, PA, HI, NH, CO, MN, FL, SC, IN, IA, TN, OR, AZ, WI, TN, IL. and the countries of: Australia, Brazil, Canada, Dominican Republic, Denmark, Ecuador, France, Germany, Israel, Kuwait, Spain, Sweden, Slovenia, Thailand, and United Kingdom.
Product codes and lots
Item Number MOL1451, Lot/Serial: 151870, 152535,152359, 152550, 152551, 152756, 152770, 160008, 160009, 160198, 160201, 160248, 160262.  Item Number MOL1455, Lot/Serial: 29230, 29232, 29233, 29453, 29454, 29669, 29670, 29671, 29689, 29690, 29691, 29692, 29845, 29846, 29847, 29848, 29849, 29850, 30271, 30272, 30273, 30274, 30275, 30609, 30610, 30611, 30612, 30613, 30615, 30616, 30617, 30619, 30620.   3M Manufacturing Lot# (Pouch and White Box of 24 DAD): 2140887, 2140332, 2127423, 2159531, 2154010, 2165240, 2173319, 2169601, 2176090, 2181518, 2181519, 2198924, 2163486, 2171337, 2176091, 2165236, 2183651, 2185129, 2216981, 2213066, 2210371, 2208204, 2205280, 2255456, 2255456, 2225778, 2230958, 2233449, 2265481 2258543.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1274-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1272-2016Class ITerminatedVoluntary: Firm initiated
Z-1273-2016Class IITerminatedVoluntary: Firm initiated
Z-1274-2016Class ITerminatedVoluntary: Firm initiated