Z-1500-2016Class II

ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD The ADVIA 560 Hematology System is...

Siemens Healthcare Diagnostics Inc. / initiated 2016-03-10 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD The ADVIA 560 Hematology System is a fully-automated, high-quality hematology system... on . FDA classified it as Class II and gives the reason as "Software anomaly; Siemens identified that software version 1.4.2133 on the ADVIA 560 Hematology System does not trigger the following flags: The G or L morphology flags for immature granulocytes (IG) and atypical...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1500-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD The ADVIA 560 Hematology System is a fully-automated, high-quality hematology system for in vitro diagnostic use in clinical laboratories
Product code
GKZ Counter, Differential Cell
Reason for recall
Software anomaly; Siemens identified that software version 1.4.2133 on the ADVIA 560 Hematology System does not trigger the following flags: The G or L morphology flags for immature granulocytes (IG) and atypical lymphocytes (ATYP), respectively. The results from patient samples which have immature granulocytes or atypical lymphocytes will not generate the flags when they should. As defined in t
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Siemens sent an Urgent Medical Device Correction Letter dated March 11, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. .Actions to be Taken by the Customer: To ensure that the patient sample amount for processing is met, follow the requirements for the amount of sample required in the ADVIA 560 Hematology System Operators Guide. Review all patient results for abnormal white blood cell results and process as defined by your lab criteria for abnormal samples and confirm using manual differentials as needed. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please review this letter with your Medical Director. Your Siemens Customer Service Engineer will contact you to schedule time for installation of the 1.4.2333 software when it is available. Please retain this letter with your laboratory records, and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative. For further questions, please call (312) 275-7795.
Quantity in commerce
US: 8 systems; Foreign: 64 systems
Distribution
Worldwide Distribution - US Distribution including Puerto Rico and to the states of : California, Illinois, Washington and Wisconsin., and to the countries of : Angola, Austria, Canada, Chile, Colombia, Czech Republic, France, Germany, India, Italy, Kenya, Netherlands, Norway, Portugal, Singapore, Slovenia, Spain, Turkey, Uganda and United Kingdom.
Product codes and lots
All serial numbers of ADVIA 560 Hematology Systems.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1500-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1500-2016Class IITerminatedVoluntary: Firm initiated