Siemens SOMATOM Definition: Intended to produce cross-sectional images of the body. Computed tomography (CT)...
Siemens Medical Solutions USA, Inc began a recall of Siemens SOMATOM Definition: Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system on . FDA classified it as Class II and gives the reason as "Software bugs in these SOMATOM systems could possibly cause scan aborts, re-scans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1518-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens SOMATOM Definition: Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Software bugs in these SOMATOM systems could possibly cause scan aborts, re-scans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens initiated a Customer Advisory Notice on 03/07/2016 to inform customers about actions for bug-fixing the following systems: SOMATOM Definition AS, SOMATOM Definition Flash and SOMATOM Definition Edge, which could possibly cause scan aborts, res cans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium. Siemens mailed certified as well as via email distribution a Customer Advisory Notice to their customers to inform them about corrective actions for bug-fixing the following systems running with software version VA44A_SP3a, VA44A_SP3b, VA44A_SP3c, correction of potential safety issues. The issues described will be corrected with software VA44_SP4, handled as update CT051/15/P. Following the installation of this update, the mentioned malfunctions will be corrected.
- Quantity in commerce
- 1096
- Distribution
- Nationwide Distribution-including DC, and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
- Product codes and lots
- Model Number: 7740769 SOMATOM Definition family of CT systems running on software versions VA44A_SP3a, VA44A_SP3b, VA44A_SP3c.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1518-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1519-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1520-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1521-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
