Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1618-2016Class II

Syva Emit Methotrexate Assay, is a homogeneous enzyme immunoassay intended for use in the quantitative...

Siemens Healthcare Diagnostics Inc. / initiated 2016-03-08 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of Syva Emit Methotrexate Assay, is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of methotrexate in human serum or plasma on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate Application (DVMTX.1) for Dimension Vista System does not meet On-Board Stability claim of 20 hours. Testing using reagent at the end of the 20 hours...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1618-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Syva Emit Methotrexate Assay, is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of methotrexate in human serum or plasma.
Product code
LAO Enzyme Immunoassay, Methotrexate
Reason for recall
Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate Application (DVMTX.1) for Dimension Vista System does not meet On-Board Stability claim of 20 hours. Testing using reagent at the end of the 20 hours On-Board Stability has shown that values of methotrexate for patient and quality control samples increased.
FDA root cause
Unknown/Undetermined by firm
Action
Siemens mailed an Urgent Medical Device Correction letter to customers informing them that they were withdrawing the Emit Methotrexate Application on the Dimension Vista System. Customers were asked to contact their local Siemens sales representative for alternative testing solutions for Methotrexate. Siemens advised customers to change the On-Board Stability for the Methotrexate application on the Dimension Vista System to four hours: alternatively, customers can run the Methotrexate application in batch mode, along with QC. It was also requested in the notification that customers complete and return the Field Correction Effectiveness Check Form that was attached with the notification within 30 days.
Quantity in commerce
18364 units
Distribution
Units were shipped to MI, FL, NJ, NY, DC, CA, GA, MI, MT, VA, NC, TX, PA, Australia, Belgium, Czech Republic, Finland, Japan, Republic of Korea, Italy, Saudi Arabia, Slovakia, Slovenia, Spain and Switzerland.
Product codes and lots
Lot # F1, F2, G1, G2, H1, and H2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1618-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1618-2016Class IITerminatedVoluntary: Firm initiated