510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1525-2016Class II

Self-Cath Pediatric Catheter Size CH 06, Product # 5044101400, Catakig # 306

MEDLINE INDUSTRIES, LP - Northfield / initiated 2016-03-22 / Terminated / root cause: manufacturing process

Coloplast Manufacturing US, LLC began a recall of Self-Cath Pediatric Catheter Size CH 06, Product # 5044101400, Catakig # 306. Product Usage: The Self-Cath intermittent urinary catheter discussed in this... on . FDA classified it as Class II and gives the reason as "Coloplast Corp. is voluntarily recalling a single lot of Self-Cath Pediatric French Size CH 06 catheters because a small portion of the lot was identified to have missing eyelets. This recall is based upon a complaint...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1525-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Self-Cath Pediatric Catheter Size CH 06, Product # 5044101400, Catakig # 306. Product Usage: The Self-Cath intermittent urinary catheter discussed in this document is a pediatric size CH06 French catheter. The catheter is indicated for use by patients with chronic urine retention and patients with a post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing the urine to drain.
Product code
EZL Catheter, Retention Type, Balloon
Reason for recall
Coloplast Corp. is voluntarily recalling a single lot of Self-Cath Pediatric French Size CH 06 catheters because a small portion of the lot was identified to have missing eyelets. This recall is based upon a complaint Coloplast received regarding a catheter malfunction.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Coloplast sent an "Urgent Medical Device Recall" letter dated March 22, 2016 to customers. The letter described the problem and the product involved in the recall. Advised consignees to immediately examine, quarantine and return the affected product. If the product has been further distributed, consignees should identify their customers and notify them of the recall. It also requested consignees to complete and return the Acknowledgement Form. For questions contact Coloplast Customer Service directly at 800-533-0464.
Quantity in commerce
31,320
Distribution
US Nationwide Distribution in the states of: AL, AR, C, FL, IL KY, LA, MD, MA, MI, MN, MO, NJ,NY, NC, OK,OR, PA, TN, TX, VA and WA.
Product codes and lots
4464131

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1525-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1525-2016Class IITerminatedVoluntary: Firm initiated