Z-1278-2016Class II
Cios Alpha, mobile X-ray system
Siemens Medical Solutions USA, Inc. / initiated 2016-02-19 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Cios Alpha, mobile X-ray system on . FDA classified it as Class II and gives the reason as "Software issues on Cios Alpha mobile C-Arm system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1278-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Cios Alpha, mobile X-ray system
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Software issues on Cios Alpha mobile C-Arm system
- Root cause tag
- software
- FDA root cause
- Software in the Use Environment
- Action
- Siemens sent an Customer Safety Advisory Notice on February 26, 2016, to inform them of some issues that may arise on the Cios Alpha unit. The letter identified the product the problem and the action needed to be taken by the customer. Users were instructed to change the date format to DDMMYYYY to prevent issues, not to release radiation as long as the DVD icon is blinking and to disconnect and re-connect the trolley if the system is needed immediately. In the meantime they are preparing a modification of Cios Alpha that will resolve the above-mentioned issues. As soon as the modification is available, a service engineer will get in touch with each customer and implement the modification to the Cios Alpha system free of charge. If the device was sold, customers were instructed to forward the recall notification to the new owner. For further questions please call (610) 448-6478.
- Quantity in commerce
- 89 systems
- Distribution
- Nationwide Distribution to the states of : AL, AR, CA, CO, CT, DE, FL, GA, IA, IL, KY, MD, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, WA and WI
- Product codes and lots
- Model # 10308191
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1278-2016 | Class II | Terminated | Voluntary: Firm initiated |