Artis One, Interventional, Fluoroscopic, x-ray system Artis one is an angiography system developed for...
Siemens Medical Solutions USA, Inc began a recall of Artis One, Interventional, Fluoroscopic, x-ray system Artis one is an angiography system developed for diagnostic imaging and interventional procedures... on . FDA classified it as Class II and gives the reason as "After importing segmentation results of the left atrium created on the Artis One system, the possibility exists for the results to be mirrored to an electro-anatomical 3D mapping system, CARTO from Biosense Webster Inc". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1282-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis One, Interventional, Fluoroscopic, x-ray system Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis One include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for patient extremities.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- After importing segmentation results of the left atrium created on the Artis One system, the possibility exists for the results to be mirrored to an electro-anatomical 3D mapping system, CARTO from Biosense Webster Inc.
- Root cause tag
- software
- FDA root cause
- Software in the Use Environment
- Action
- Siemens mailed an Important Customer Safety Notice to customers to inform them of the issue and the corrective actions they plan to take to address the problem. The letter stated that Siemens service organization will contact customers to arrange a date for the installation of the software update. Customers were instructed to call Siemens service organization for an appointment at 1-800-888-7436.
- Quantity in commerce
- 1 angiography system
- Distribution
- Nationwide Distribution to Michigan only
- Product codes and lots
- Serial # 82010, material # 10848600
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1282-2016 | Class II | Terminated | Voluntary: Firm initiated |