TC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 8 LEFT, STERILE R, REF 12 000 005, S&N 7500260...
Smith & Nephew, Inc. began a recall of TC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 8 LEFT, STERILE R, REF 12 000 005, S&N 7500260, Smith & Nephew Orthopaedics AG Product Usage: Tibial... on . FDA classified it as Class II and gives the reason as "The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1538-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- TC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 8 LEFT, STERILE R, REF 12 000 005, S&N 7500260, Smith & Nephew Orthopaedics AG Product Usage: Tibial knee prosthesis
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm, Smith & Nephew, issued an "URGENT FIELD SAFETY NOTICE: Medical Device Field Safety Corrective Action/Recall" letter dated 3/10/2016 to its local QA/RA personnel/customers in the affected countries by email on 03/11/2016. The letter described the product, problem and actions to be taken. The customers were instructed to locate and quarantine affected unused devices immediately; return quarantined product to your national Smith & Nephew agency/distributor; complete and return the Return Slip via fax to your national Smith & Nephew agency/distributor; and pass on this safety information to all those who need to be aware in your organization. The local QA/RA staff will then notify end customers according to local procedures. In Germany as well as in Switzerland, sales representatives will inform the affected customers in-person and retrieve affected devices from their stock. If you have any questions, call 978-749-1330 or email: [email protected]
- Quantity in commerce
- 6 units
- Distribution
- International Distribution only in countries of: Germany and Switzerland.
- Product codes and lots
- Batch number C1507158
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1537-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1538-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1539-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1540-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1541-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1542-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1543-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1544-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1545-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1546-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1547-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1548-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1549-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1550-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1551-2016 | Class II | Terminated | Voluntary: Firm initiated |