Z-1302-2016Class II

Siemens branded SOMATOM Emotion and SOMATOM Sensation families of CT systems intended to produce...

Siemens Medical Solutions USA, Inc. / initiated 2016-02-15 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens branded SOMATOM Emotion and SOMATOM Sensation families of CT systems intended to produce cross-sectional images of the body by computer reconstruction... on . FDA classified it as Class II and gives the reason as "Siemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1302-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens branded SOMATOM Emotion and SOMATOM Sensation families of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Siemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio licenses "Cardio Pro" and "Cardio Retro".
Root cause tag
design
FDA root cause
Device Design
Action
A Safety Advisory Notice was sent to customers informing them of a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems. Siemens informed them that they will deliver an Update Instruction for their Service Engineers to modify the installed Firmware. This will correct the rare synchronization errors of Scanner Firmware and Scanner Software to start the application of the x-ray in compliance with the optimum triggering point for best image quality.
Quantity in commerce
124 units
Distribution
US (nationwide) including Guam and the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, VA, and WA.
Product codes and lots
Model Number: SOMATOM Emotion Duo (2003) - 3813933

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1302-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1300-2016Class IITerminatedVoluntary: Firm initiated
Z-1301-2016Class IITerminatedVoluntary: Firm initiated
Z-1302-2016Class IITerminatedVoluntary: Firm initiated
Z-1303-2016Class IITerminatedVoluntary: Firm initiated
Z-1304-2016Class IITerminatedVoluntary: Firm initiated
Z-1305-2016Class IITerminatedVoluntary: Firm initiated
Z-1306-2016Class IITerminatedVoluntary: Firm initiated
Z-1307-2016Class IITerminatedVoluntary: Firm initiated