Z-1569-2016Class II

enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V...

Ortho-Clinical Diagnostics, Inc. / initiated 2016-03-17 / Terminated / root cause: design

Ortho-Clinical Diagnostics began a recall of enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL NAME: enGen (TM) Track System -- NOTE... on . FDA classified it as Class II and gives the reason as "The electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may loosen from its correct position above the vibration feeder support plate and migrate to under the support plate. If this...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1569-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL NAME: enGen (TM) Track System -- NOTE: Outside the US the system is also known as VITROS Automation Solution (VAS) TC Automation System (configured with Thermo Scientific Recapper Module from VITROS Automation Solutions), Product Code 6844300 --- The enGen (TM) Lab Automation System is a modification to VITROS Analyzers that includes the addition of an automated track sub-system (TC Automation) for pre-analytical and post-analytical sample and data management. In addition to an Automated Track Sub-system, the enGen (TM) System includes one or more VITROS Chemistry, Immunodiagnostic and/or Integrated Systems (clinical analyzers) as well as a Data Management Subsystem (Middleware), and a Software Configuration (gsb) File. The enGen System is typically interfaced with a Laboratory Information System (LIS) via connectivity facilitated at the Data Management Subsystem level. The enGen (TM) Laboratory Automation System contains a module (i.e. enGen (TM)Laboratory Automation Recapper Module) that applies caps to the sample collection container after the sample has been aspirated. In this module, an electrical cable supplies power to the cap vibratory feeder and is supposed to be located above the vibratory feeder support plate. The cap vibratory feeder sits upon a feeder support plate. Un-capped sample containers pass underneath this support plate prior to being re-capped for storage. The Recapper Module is supplied by Thermo Fisher Scientific.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may loosen from its correct position above the vibration feeder support plate and migrate to under the support plate. If this occurs, the uncapped sample tubes may hit the cable, and sample fluid (e.g., serum) may splash from the tube and into another tube or onto the under carriage of the support plate.
Root cause tag
design
FDA root cause
Component design/selection
Action
On 3/17/2016 URGENT PRODUCT CORRECTION NOTIFICATION letter (Ref. CL2016-074, dated 3/17/2016)) was sent notifying the users of the issue and advise them of the required actions. The notification was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address) to all customers who received the affected product and advised them of the issue and the required actions. Foreign affiliates were informed by email on 3/17/2016 of the issue and instructed to notify their consignees of the issue and required actions.
Quantity in commerce
US: 19, Foreign: 30
Distribution
Distributed in the states of MD, NY, ID, NC, CO, OH, PA, TN, MS, MO, CA, DE, FL, GA, AZ, NJ, and WV and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, France, India, Italy, Norway, Portugal, Saudi Arabia, Spain, Sweden, Thailand, and United Kingdom.
Product codes and lots
US Serial Numbers: JFBX395J, GBX395J, J953252, JJXBQW4J, J3RHC23J, J86PWC5J, J5LQ6Q2J, JFCM073J, J54FC01J, J5J4B45J, J7KSC15J, J4HX7S4J, JGWXQ52J, J953221, JG7BFB5J, JHPKG25J, JCXL941J, J736BM4J, JFXCBG3J; -- Foreign Serial Numbers: J89133 (VAS), JHWJZY4J, J3J4B45J, J953244, JCT20Q3J, J16XQH4J, GWJVY4J, DWX052J,  J271264, J271368, J271371, J271269, J85206, J953216, DP2LF62, J7023874, J7023847, J7025017, J7024893, J7023645, J7023435, J57663, J800103, S/N 89VQ762 (VAS), J75863, J75691, J30372, J3055503, J3055496.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1569-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1569-2016Class IITerminatedVoluntary: Firm initiated