Z-1519-2016Class II

Siemens SOMATOM Definition AS; Intended to produce cross-sectional images of the body. Computed tomography...

Siemens Medical Solutions USA, Inc. / initiated 2016-03-02 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens SOMATOM Definition AS; Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system on . FDA classified it as Class II and gives the reason as "Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1519-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens SOMATOM Definition AS; Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.
Root cause tag
software
FDA root cause
Software design
Action
Siemens initiated a Customer Advisory Notice on 03/07/2016 to inform customers about actions for bug-fixing the following systems: SOMATOM Definition AS, SOMATOM Definition Flash and SOMATOM Definition Edge, which could possibly cause scan aborts, res cans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium. Siemens mailed certified as well as via email distribution a Customer Advisory Notice to their customers to inform them about corrective actions for bug-fixing the following systems running with software version VA44A_SP3a, VA44A_SP3b, VA44A_SP3c, correction of potential safety issues. The issues described will be corrected with software VA44_SP4, handled as update CT051/15/P. Following the installation of this update, the mentioned malfunctions will be corrected.
Quantity in commerce
1096
Distribution
Nationwide Distribution-including DC, and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Product codes and lots
Model Number: 8098027 SOMATOM Definition family of CT systems running on software versions VA44A_SP3a, VA44A_SP3b, VA44A_SP3c.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1519-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1518-2016Class IITerminatedVoluntary: Firm initiated
Z-1519-2016Class IITerminatedVoluntary: Firm initiated
Z-1520-2016Class IITerminatedVoluntary: Firm initiated
Z-1521-2016Class IITerminatedVoluntary: Firm initiated