SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images of the body by...
Siemens Medical Solutions USA, Inc began a recall of SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission dta from... on . FDA classified it as Class II and gives the reason as "Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1350-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission dta from either the same axial plane taken at different angles or spiral planes taken at different angles.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened during service. This could not happen during clinical operation. Therefore, there is no potential risk for the user or patient and the CT System has no risk for a malfunction during clinical operation
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Siemens sent an "Informational Only" notice to customers on 03/22/2016 informing them to be aware that the front cover on some Siemens CT systems was not properly welded. The letter was only to notify customers that the issue is being addressed with the completion of a service visit. In addition, Siemen plans to replace each cover and hand delivered each notification to customers at the time the service is performed.
- Quantity in commerce
- 1 CT system
- Distribution
- Distributed to: Cincinnati, OH.
- Product codes and lots
- Model Number: SOMATOM Definition AS-8098027 Serial # SOMATOM Definition AS- 95766
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1350-2016 | Class II | Terminated | Voluntary: Firm initiated |